FDA Adverse Event Injury Summary report: N

BI METRIC/XR LAT RPP C/L 10MM

MDR report key: 7440759 · Received April 18, 2018

Report

Report Number
0001825034-2018-02851
Event Type
Injury
Date Received
April 18, 2018
Date of Event
January 16, 2017
Report Date
April 19, 2018
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
PK052089
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

THE FOLLOWING SECTIONS COULD NOT BE COMPLETED WITH THE LIMITED INFORMATION PROVIDED. REPORT SOURCE: (B)(6). CONCOMITANT PRODUCT(S): A 11-163679 28MM M2A HI CARBON HD +3MM NK 902340; 15-105011 M2A TPR HI CARBON 37/28MM LNR 841770; 103531 TI LOW PROFILE SCREW 6.5X20MM 845130; 103531 TI LOW PROFILE SCREW 6.5X20MM 845150; 10-104050 M2A-T M/H RAD 2HL SHL 37/50MM 587360; X11-180440 BI METRIC/XR LAT RPP C/L 10MM 607080; REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS PROVIDED. DEDICATED IMAGES 9 YEARS AFTER PRIMARY OPERATION OF THE RIGHT HIP CONFIRMS A ZONE OF LUCENCY SURROUNDING THE ACETABULAR SCREWS. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

PATIENT¿S RIGHT HIP WAS REVISED APPROXIMATELY 10 YEARS POST-IMPLANTATION DUE TO PAIN AND EVIDENCE OF A PSEUDOTUMOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281800 BI METRIC/XR LAT RPP C/L 10MM HIP PROSTHESIS LPH ZIMMER BIOMET, INC. N/A 607080

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| R