HUDSON CIRCUIT,NEONATAL,DUAL HTD LIMB W/18 IN
Report
- Report Number
- 3004365956-2018-00115
- Event Type
- Malfunction
- Date Received
- April 17, 2018
- Date of Event
- April 18, 2018
- Report Date
- April 3, 2018
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- BZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PER MEDWATCH REPORT: FEMALE BABY (B)(6); DIAGNOSED WITH PREMATURITY AND RESPIRATORY DISTRESS ON CPAP. DATE OF EVENT CORRECTED TO (B)(6) 2018. ONE INSPIRATORY LEG OF THE INFANT BREATHING CIRCUIT WAS RECEIVED FOR INVESTIGATION. A VISUAL EXAM WAS PERFORMED AND IT WAS OBSERVED THAT THE CLEAR PLASTIC TEMPERATURE PROBE PORTAL WAS BROKEN. BASED ON A PRELIMINARY INVESTIGATION, THE BREAK IN THE TEMPERATURE PROBE PORTAL IS AT A SONIC WELD. THE PROBABLE CAUSE IS UNKNOWN AT THIS TIME. A CAPA WAS OPENED TO ADDRESS THIS ISSUE. A CONCLUSION CODE COULD NOT BE FOUND AS THE COMPLAINT WAS CONFIRMED; HOWEVER, A ROOT CAUSE COULD NOT BE ESTABLISHED.
(B)(4). THE ACTUAL DEVICE INVOLVED IN THIS COMPLAINT WAS NOT RETURNED FOR EVALUATION. TWO PHOTOS OF CATALOG NUMBER 780-07 WERE RECEIVED FOR ANALYSIS. IT CAN BE OBSERVED THE TEE ADAPTOR IP-6234 BROKEN AS DESCRIBED IN THE COMPLAINT DESCRIPTION. NO OTHER ISSUES WERE OBSERVED. THE DEVICE HISTORY RECORD OF BATCH NUMBER 74D1700921 HAS BEEN REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND WHICH COULD POTENTIALLY RELATE TO THIS COMPLAINT. ALSO, THE DHR OF (B)(4) WITH LOT NUMBER 2219459 WAS REVIEWED AND ONE PARAMETER WAS FOUND OUT OF TOLERANCE ACCORDING TO PARAMETER SHEET (B)(4), REV.00 FOR ULTRASONIC WELDER MACHINE. CUSTOMER COMPLAINT IS CONFIRMED BASED ON THE VISUAL INSPECTION OF THE PHOTOGRAPHS SUBMITTED. ROOT CAUSE IS A PARAMETER OUT OF TOLERANCE. AN NC WAS OPENED TO FURTHER ADDRESS THE ISSUE OF TEE ADAPTOR IP-6234.
CUSTOMER COMPLAINT ALLEGES "THE CIRCUIT CONNECTION WHERE IT CONNECTS TO THE REMOTE TUBING FOR THE ISOLETTE CRACKED AND BROKE ." ALLEGED EVENT REPORTED AS OCCURRED DURING USE. IT WAS REPORTED THERE WAS NO INJURY OR CONSEQUENCE TO THE PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE".
CUSTOMER COMPLAINT ALLEGES "THE CIRCUIT CONNECTION WHERE IT CONNECTS TO THE REMOTE TUBING FOR THE ISOLETTE CRACKED AND BROKE ." ALLEGED EVENT REPORTED AS OCCURRED DURING USE. IT WAS REPORTED THERE WAS NO INJURY OR CONSEQUENCE TO THE PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279894 | HUDSON CIRCUIT,NEONATAL,DUAL HTD LIMB W/18 IN | SET, TUBING AND SUPPORT,VENTILATOR | BZO | TELEFLEX MEDICAL | 74D1700921 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 DA |