FDA Adverse Event Malfunction Summary report: N

BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE

MDR report key: 7436850 · Received April 17, 2018

Report

Report Number
1213809-2018-00237
Event Type
Malfunction
Date Received
April 17, 2018
Date of Event
April 3, 2018
Report Date
May 10, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
MEG
UDI-DI
30382903052708
PMA / PMN Number
K011103
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT, THEREFORE A ROOT CAUSE COULD NOT BE DETERMINED. DHR REVIEW FOR BATCH 6201984 (P/N 305270): MANUFACTURING DATES: 07/31/2016 TO 08/08/2016. BATCH QUANTITY WAS (B)(4). ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 6201984 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. THE REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. ROOT CAUSE AND CAPA NOT REQUIRED AS NO DEFECTS WERE CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE LEAKED FROM THE BOTTOM WHEN THE PLUNGER WAS PULLED DOWN. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281083 BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE SAFETY ENGINEERED SYRINGE AND NEEDLE MEG BECTON DICKINSON MEDICAL SYSTEMS 6201984 30382903052708

Patients

Seq Age Sex Outcome Treatment
1 Other