FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 7435926 · Received April 17, 2018

Report

Report Number
1823260-2018-01186
Event Type
Malfunction
Date Received
April 17, 2018
Date of Event
March 17, 2018
Report Date
May 17, 2018
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER RETURNED METER AND STRIPS FOR INVESTIGATION. THE RETURNED METER AND STRIPS WERE TESTED IN COMPARISON TO A REFERENCE METER AND MASTER LOT STRIPS. TWO HUMAN BLOOD SAMPLES FROM WARFARIN DONORS AND INTERNAL REFERENCE METERS WERE USED. TESTING RESULTS: MARCUMAR 3: REFERENCE METER WITH MASTERLOT STRIPS: 2.8 INR, CUSTOMER METER WITH CUSTOMER STRIPS: 2.6 INR. MARCUMAR 4:: REFERENCE METER WITH MASTERLOT STRIPS: 3.4 INR, CUSTOMER METER WITH CUSTOMER STRIPS: 3.1 INR. THE MAXIMUM DIFFERENCE BETWEEN MEASUREMENTS WITH THE SAME BLOOD SAMPLE WAS 9 %. RETURNED CUSTOMER MATERIAL AND RETENTION MATERIAL COMPLIES WITH SPECIFICATION.

Additional Manufacturer Narrative · 1

UNIQUE IDENTIFIER (UDI)#: (B)(4). THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

THE CUSTOMER'S SON STATED THEY RECEIVED QUESTIONABLE RESULTS FROM COAGUCHEK XS METER SERIAL NUMBER (B)(4). THE RESULT AT 10:10 AM WAS 4.3 INR AND THE RESULT AT 10:18 AM WAS 2.6 INR. THE CUSTOMER USED DIFFERENT FINGERS FOR THE TESTS. THE CUSTOMER WAS NOT HARMED AND DID NOT GO TO THE HOSPITAL. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER HAD NO HEPARIN. THE HEMATOCRIT WAS NOT KNOWN. THE THERAPEUTIC RANGE WAS 2.5-3.5 INR. THE SUSPECT METER AND STRIPS WERE REQUESTED TO BE RETURNED FOR INVESTIGATION. RELEVANT RETENTION TEST STRIPS (LOT 244033) WERE TESTED IN COMPARISON WITH THE MASTER LOT COAGUCHEK XS PT. FOR THIS PURPOSE, TWO HUMAN BLOOD SAMPLES FROM MARCUMAR DONORS AND TWO INTERNAL REFERENCE METERS WERE USED. RETENTION SAMPLES WERE ACCEPTABLE. NO ERROR MESSAGES OCCURRED AND RETENTION MATERIAL COMPLIED WITH SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279112 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 24403311

Patients

Seq Age Sex Outcome Treatment
1 87 YR MARCUMAR| METOPROLOL| RAMIPRIL| THORASEMIDE