FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® TRACE ELEMENT SERUM BLOOD COLLECTION TUBES

MDR report key: 7434048 · Received April 16, 2018

Report

Report Number
1917413-2018-00723
Event Type
Malfunction
Date Received
April 16, 2018
Date of Event
October 18, 2016
Report Date
April 11, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903683814
PMA / PMN Number
K041071
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR 6090562 WITH THE INCIDENT LOT WAS OBSERVED. ROOT CAUSE DESCRIPTION: THE FOLLOWING INFORMATION EXPLAINS THE APPEARANCE OF THE PATTERN WITHIN THE TUBES: THE ADDITIVE FOR THE CAT NO PRODUCT IS APPLIED ONTO THE INTERIOR WALLS OF THE BLOOD COLLECTION VESSEL VIA THE SPRAYCOATING OF A PRESCRIBED AMOUNT OF WATER BASED K2EDTA SOLUTION. THE WATER IS THEN DRIED, LEAVING THE CHARACTERISTIC ¿MIST¿ DOT PATTERN OF RESIDUAL K2EDTA ON THE VESSEL WALL. IN RARE INSTANCES WHERE ASSEMBLY MACHINE MISALIGNMENTS OCCUR, ADDITIVE DISPENSE NOZZLES MAY NOT BE CENTERED WITHIN THE TUBES¿ OPENING AND THE NOZZLES¿ POSITION IS BIASED MORE TOWARDS ONE OF THE VESSEL¿S HEMISPHERES. IN SUCH INSTANCES, ONE SIDE (OR HEMISPHERE) OF THE TUBE VESSEL INTERIOR RECEIVED A HEAVIER COATING OF ADDITIVE SOLUTION, WHILE POINTS ON THE OPPOSING SIDE OF VESSEL RECEIVE A LIGHTER MISTING. LAGER DROPLETS OF THE SOLUTION IN CLOSE PROXIMITY TO ONE ANOTHER THEN TEND TO COALESCE, AND ONCE THE WATER IS DRIED, LEAVE BEHIND A WAFER LIKE DEPOSIT OF THE ADDITIVE. THIS ADDITIVE DEPOSIT VISUALLY STANDS OUT AS IT DOES NOT APPEAR LIKE THE TYPICAL DOT PATTERN FOR THIS TUBE TYPE. INSPECTIONS ARE PERFORMED IN PROCESS AND AT FINAL INSPECT AT A PRESCRIBED RATE AS SPECIFIED IN THE PRODUCT SPECIFICATION. THE VISION INSPECTION ONLINE PERFORMS A CHECK FOR PRESENCE OF ADDITIVE AND IS NOT QUANTITATIVE. THERE HAS BEEN TRAINING THAT HAS BEEN ROLLED OUT TO ALL OPERATORS ALONG WITH A JOB AID FOR UPKEEP OF THE SPRAY COAT THAT WILL HELP WITH THE PREVENTION OF THE DROPLETS. ALONG WITH THE TRAINING THE BED CHAIN WAS CHANGED OUT. THIS WILL HELP IN ASSURING THAT THE TUBES GET PROPERLY ALIGNED WHEN GOING INTO THE SPRAY COAT.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® TRACE ELEMENT SERUM BLOOD COLLECTION TUBES HAD FOREIGN MATTER IN TUBE. NO SERIOUS INJURY. NO MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277943 BD VACUTAINER® TRACE ELEMENT SERUM BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. 6090562 50382903683814

Patients

Seq Age Sex Outcome Treatment
1 Other