TECNIS ITEC PRELOADED 1-PIECE IOL
Report
- Report Number
- 2648035-2018-00527
- Event Type
- Malfunction
- Date Received
- April 14, 2018
- Date of Event
- March 14, 2018
- Report Date
- November 12, 2018
- Manufacturer
- JOHNSON AND JOHNSON SURGICAL VISION, INC.
- Product Code
- HQL
- UDI-DI
- 05050474558311
- PMA / PMN Number
- P980040
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHARMACIST
Narratives
DEVICE EVALUATION: THE DEVICE WAS NOT RETURNED AT THE MANUFACTURING SITE; THEREFORE PRODUCT TESTING COULD NOT BE PERFORMED AND THE CUSTOMER'S REPORTED COMPLAINT COULD NOT BE VERIFIED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE INTRAOCULAR LENS (IOL) WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. HISTORICAL DATA ANALYSIS: A SEARCH REVEALED THAT NO ADDITIONAL INVESTIGATION REQUESTS FOR THIS ORDER NUMBER HAVE BEEN RECEIVED. LABELING REVIEW: THE DIRECTIONS FOR USE (DFU) WERE REVIEWED. THE DIRECTIONS FOR USE (DFU) ADEQUATELY PROVIDES INSTRUCTIONS AND PRECAUTIONS ALONG WITH WARNINGS FOR THE PROPER USE AND HANDLING OF THE DEVICE. AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY AND THE REPORTED ISSUE COULD NOT BE VERIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IF IMPLANTED, GIVE DATE: NOT APPLICABLE AS THE LENS WAS NOT IMPLANTED. IF EXPLANTED, GIVE DATE: NOT APPLICABLE AS THE LENS WAS NOT IMPLANTED AND THEREFORE NOT EXPLANTED (B)(6). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING THE IMPLANTATION OF AN INTRAOCULAR LENS (MODEL PCB00) THE PLUNGER WAS JAMMED PREVENTING THE FULL LENS DELIVERY. REPORTEDLY, THE LENS HAD CONTACT WITH THE EYE BUT THERE WAS NO HARM FOR THE PATIENT. ANOTHER LENS WAS IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274424 | TECNIS ITEC PRELOADED 1-PIECE IOL | MONOFOCAL IOLS | HQL | JOHNSON AND JOHNSON SURGICAL VISION, INC. | PCB00 | 05050474558311 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |