FDA Adverse Event Malfunction Summary report: N

TECNIS ITEC PRELOADED 1-PIECE IOL

MDR report key: 7431644 · Received April 14, 2018

Report

Report Number
2648035-2018-00527
Event Type
Malfunction
Date Received
April 14, 2018
Date of Event
March 14, 2018
Report Date
November 12, 2018
Manufacturer
JOHNSON AND JOHNSON SURGICAL VISION, INC.
Product Code
HQL
UDI-DI
05050474558311
PMA / PMN Number
P980040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE WAS NOT RETURNED AT THE MANUFACTURING SITE; THEREFORE PRODUCT TESTING COULD NOT BE PERFORMED AND THE CUSTOMER'S REPORTED COMPLAINT COULD NOT BE VERIFIED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE INTRAOCULAR LENS (IOL) WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. HISTORICAL DATA ANALYSIS: A SEARCH REVEALED THAT NO ADDITIONAL INVESTIGATION REQUESTS FOR THIS ORDER NUMBER HAVE BEEN RECEIVED. LABELING REVIEW: THE DIRECTIONS FOR USE (DFU) WERE REVIEWED. THE DIRECTIONS FOR USE (DFU) ADEQUATELY PROVIDES INSTRUCTIONS AND PRECAUTIONS ALONG WITH WARNINGS FOR THE PROPER USE AND HANDLING OF THE DEVICE. AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY AND THE REPORTED ISSUE COULD NOT BE VERIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

IF IMPLANTED, GIVE DATE: NOT APPLICABLE AS THE LENS WAS NOT IMPLANTED. IF EXPLANTED, GIVE DATE: NOT APPLICABLE AS THE LENS WAS NOT IMPLANTED AND THEREFORE NOT EXPLANTED (B)(6). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANTATION OF AN INTRAOCULAR LENS (MODEL PCB00) THE PLUNGER WAS JAMMED PREVENTING THE FULL LENS DELIVERY. REPORTEDLY, THE LENS HAD CONTACT WITH THE EYE BUT THERE WAS NO HARM FOR THE PATIENT. ANOTHER LENS WAS IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274424 TECNIS ITEC PRELOADED 1-PIECE IOL MONOFOCAL IOLS HQL JOHNSON AND JOHNSON SURGICAL VISION, INC. PCB00 05050474558311

Patients

Seq Age Sex Outcome Treatment
1