FISHCER CONE BIOPSY EXCISOR
Report
- Report Number
- 1221923-2006-00002
- Event Type
- Malfunction
- Date Received
- July 21, 2006
- Date of Event
- May 18, 2006
- Report Date
- July 20, 2006
- Manufacturer
- APPLE MEDICAL CORP.
- Product Code
- KNW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- OTHER
Narratives
THE FCBE DESIGN HAS BEEN VALIDATED BY TESTING TO THE 2 STANDARDS THAT COVER ELECTROSURGICAL UNITS (ESUS) AND THEIR ACCESSORIES; ANSI/AAMI HF18 AND IEC 60601-2-2. IN ADDITION, EACH FCBE UNDERGOES A TOP ASSEMBLY VALIDATION HYPOT TEST. THE DIRECTIONS FOR USE (DFU) FOR THE FCBE PRESCRIBES ESU SETTINGS THAT ARE BASED ON DEVICE DESIGN. HOWEVER, IF THE FCBE IS OPERATED UNDER CONDITIONS THAT EXCEED THE DESIGN BASIS, OR BY USERS NOT ADEQUATELY TRAINED, THE FCBE MAY FAIL IN A MANNER CONSISTENT WITH THIS REPORTED FAILURE. OPERATION BEYOND THE DESIGN BASIS MAY BE INTENTIONAL OR UNINTENTIONAL DUE TO A FAULTY OR OUT OF CALIBRATION ESU. THE CIRCUMSTANCES OF THIS EVENT HOWEVER SUGGEST THAT INEXPERIENCE IN THE USE OF THE FCBE DEVICES MAY HAVE CONTRIBUTED TO THE EVENT. APPLE MEDICAL WILL IN-SERVICE THE FANNIN FACILITY IN THE PROPER USE OF THE FCBE.
CARMEL OF FANNIN SURGICARE CALLED APPLE MEDICAL TO REPORT THE MALFUNCTION OF 3 FISCHER CONE BIOPSY EXCISROS (FCBE'S), A DEVICE MFR'D BY APPLE. CARMEL REPORTED THAT "WE'VE HAD THE WIRES BREAK ON 2 OF THE FISCHER PRODUCT" WHILE IN USE DURING A PROCEDURE. ONE FCBE WAS ALSO REPORTED TO BE "STICKING IN THE PENCIL". SHE REPORTED THAT THEIR "EXPERIENCED USER", DR. IRWIN, HAS HAD NO PROBLEMS. BUT NEW USERS ARE EXPERIENCING PROBLEMS, LIKE WIRES BREAKING. CARMEL STATED THAT A VALLEYLAB FORCE-TX ESU WAS BEING USED AT A SETTING OF 50 WATTS IN "CUT" MODE. CARMEL REPORTED THAT NO PT INJURIES HAVE OCCURRED AND THAT THERE WAS NO PART OR PIECE OF THE FAILED ELECTRODE LEFT IN THE PT. AN APPLE MEDICAL SALES REP WILL BE DISPATCHED TO THE FANNIN FACILITY TO CONTINUE THE INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FISHCER CONE BIOPSY EXCISOR | BIOPSY INSTRUMENT | KNW | APPLE MEDICAL CORP. | MEDIUM/LARGE SHALLO | I265L/H235O |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |