FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE

MDR report key: 7419480 · Received April 10, 2018

Report

Report Number
1920898-2018-00216
Event Type
Malfunction
Date Received
April 10, 2018
Date of Event
March 27, 2018
Report Date
May 11, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00382903249107
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CUSTOMER RETURNED ONE LOOSE 3/10CC, 6MM SYRINGE. THE RETURNED SYRINGE WAS EXAMINED AND EXHIBITED HARD, CLEAR MATERIAL ON THE SURFACE OF THE CANNULA SHAFT. A SMALL PORTION OF THIS MATERIAL WAS REMOVED FROM THE SAMPLE AND PREPARED FOR FTIR SPECTRAL ANALYSIS. THE SPECTRAL ANALYSIS SUGGESTS THAT THIS MATERIAL HAS COMPONENTS SIMILAR TO THOSE OF ADHESIVE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7023681 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (ADHESIVE SPLATTER ON CANNULA). PROBABLE ROOT CAUSE DETERMINED TO BE MISALIGNMENT DURING APPLICATION OF ADHESIVE ON THE NEEDLE LINES. WHEN THIS OCCURS, ADHESIVE RUNOVER ONTO THE HUB OR POSSIBLE THE CANNULA MAY OCCUR. ADDITIONALLY, ADHESIVE MAY BE INACCURATELY APPLIED TO THE RUBBERIZED PULL WHEEL ON THE LINE, WHICH CAN ADDITIONALLY TRANSFER ADHESIVE TO THE CANNULA DURING ROUTINE USE. CAPA (B)(4) WAS INITIATED BY THE HOLDREGE PLANT TO ADDRESS ADHESIVE RUNOVER.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND ON THE SURFACE OF A BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE WHILE PREPARING FOR INJECTION. THE DEVICE WAS NOT USED AND THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253506 BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE INSULIN SYRINGE AND NEEDLE FMF BD MEDICAL - DIABETES CARE 7023681 00382903249107

Patients

Seq Age Sex Outcome Treatment
1 Other