FDA Adverse Event Injury Summary report: N

BAXTER INTERNATIONAL INC.

MDR report key: 741812 · Received July 10, 2006

Report

Report Number
741812
Event Type
Injury
Date Received
July 10, 2006
Date of Event
June 5, 2006
Report Date
June 13, 2006
Manufacturer
BAXTER INTERNATIONAL INC.
Product Code
FRN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MD
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

4 GM MSO4 PROGRAMMED INTO IV PUMP CHAMBER "C" AT 100CC/HR. PT DEVELOPED SUDDEN AND SEVERE HYPOTENSION. IT WAS NOTED THAT ENTIRE BAG OF MGSO4 HAD BEEN BOLUSED IN LESS THAN 5 MIN. IV PUMP CONTINUED TO READ THAT IT WAS RUNNING AT 100CC/HR. PT REQUIRED THE ADMINISTRATION OF DOPAMINE AND 1000CC IV FLUID BOLUS. BP RECOVERED WITHIN 30 MINUTES WITHOUT RESIDUAL EFFECT. THIS UNIT WAS A RENTAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BAXTER INTERNATIONAL INC. IV PUMP FRN BAXTER INTERNATIONAL INC. COLLEAGUE 3 *

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention