SAGITTAL SAW ATTACHMENT
Report
- Report Number
- 8030965-2018-52951
- Event Type
- Malfunction
- Date Received
- April 10, 2018
- Date of Event
- February 23, 2018
- Report Date
- March 16, 2018
- Manufacturer
- DEPUY SYNTHES PRODUCTS LLC
- Product Code
- HWE
- UDI-DI
- 7611819408319
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. D1, D2, D2B, D4: THE CATALOG NUMBER WAS REPORTED AS 532.101 IN THE INITIAL REPORT AND HAS BEEN UPDATED TO 532.021. THE DEVICE PRODUCT CODE, COMMON DEVICE NAME AND UDI HAVE BEEN UPDATED ACCORDINGLY. G1: THE MANUFACTURING SITE NAME WAS DOCUMENTED AS DEPUY SYNTHES PRODUCTS LLC (OBERDORF) IN THE INITIAL REPORT. THIS HAS BEEN UPDATED TO SYNTHES PRODUKTIONS GMBH (WALDENBURG). PLEASE NOTE THAT THE CONTACT OFFICE NAME/ADDRESS HAVE BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECTED MANUFACTURING FACILITY. H4: THE DEVICE MANUFACTURE DATE WAS DOCUMENTED AS JULY 5, 2017 IN THE INITIAL REPORT AND HAS BEEN UPDATED TO MARCH 27, 2009. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE DEVICE BEARING WAS NOT FUNCTIONING AND DEFECTIVE AND THE DEVICE FAILED PRE-TEST FOR VERIFY IF SECURED WITH PIN. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER MAINTENANCE, WHICH IS USER ERROR/MISUSE/ABUSE. I F ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE AND THE ATTACHMENT DEVICE WHILE BEING USED TOGETHER, WERE LEAKING OIL. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256721 | SAGITTAL SAW ATTACHMENT | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | DEPUY SYNTHES PRODUCTS LLC | 7611819408319 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |