FDA Adverse Event Malfunction Summary report: N

SAGITTAL SAW ATTACHMENT

MDR report key: 7415007 · Received April 10, 2018

Report

Report Number
8030965-2018-52951
Event Type
Malfunction
Date Received
April 10, 2018
Date of Event
February 23, 2018
Report Date
March 16, 2018
Manufacturer
DEPUY SYNTHES PRODUCTS LLC
Product Code
HWE
UDI-DI
7611819408319
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. D1, D2, D2B, D4: THE CATALOG NUMBER WAS REPORTED AS 532.101 IN THE INITIAL REPORT AND HAS BEEN UPDATED TO 532.021. THE DEVICE PRODUCT CODE, COMMON DEVICE NAME AND UDI HAVE BEEN UPDATED ACCORDINGLY. G1: THE MANUFACTURING SITE NAME WAS DOCUMENTED AS DEPUY SYNTHES PRODUCTS LLC (OBERDORF) IN THE INITIAL REPORT. THIS HAS BEEN UPDATED TO SYNTHES PRODUKTIONS GMBH (WALDENBURG). PLEASE NOTE THAT THE CONTACT OFFICE NAME/ADDRESS HAVE BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECTED MANUFACTURING FACILITY. H4: THE DEVICE MANUFACTURE DATE WAS DOCUMENTED AS JULY 5, 2017 IN THE INITIAL REPORT AND HAS BEEN UPDATED TO MARCH 27, 2009. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE DEVICE BEARING WAS NOT FUNCTIONING AND DEFECTIVE AND THE DEVICE FAILED PRE-TEST FOR VERIFY IF SECURED WITH PIN. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER MAINTENANCE, WHICH IS USER ERROR/MISUSE/ABUSE. I F ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE AND THE ATTACHMENT DEVICE WHILE BEING USED TOGETHER, WERE LEAKING OIL. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256721 SAGITTAL SAW ATTACHMENT INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES PRODUCTS LLC 7611819408319

Patients

Seq Age Sex Outcome Treatment
1