HEARTWARE VENTRICULAR ASSIST SYSTEM - APEX CORING TOOL
Report
- Report Number
- 3007042319-2018-01391
- Event Type
- Malfunction
- Date Received
- April 10, 2018
- Date of Event
- March 15, 2018
- Report Date
- November 1, 2018
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- UDI-DI
- 00888707000154
- PMA / PMN Number
- P100047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: AN APICAL CORING TOOL ASSOCIATED WITH SURGICAL TOOL KIT (LOT# 1377569) WAS RETURNED FOR ANALYSIS. REVIEW OF THE MANUFACTURING DOCUMENTATION CONFIRMED THAT THE ASSOCIATED DEVICE MET ALL REQUIREMENTS FOR RELEASE. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE SURGICAL TOOL IN RELATION TO THE REPORTED EVENT. VISUAL EXAMINATION OF THE APICAL CORING TOOL REVEALED NO OBVIOUS PHYSICAL DAMAGE OR ANOMALIES. FUNCTIONAL ANALYSIS DEMONSTRATES THAT THE DEVICE EXTENDS AND RETRACTS AS INTENDED. AS A RESULT, THE REPORTED EVENT OF CORING TOOL ¿NOT CUTTING¿ WAS NOT CONFIRMED. BASED ON THE INFORMATION AVAILABLE, THERE IS NO EVIDENCE TO INDICATE THAT A DEVICE MALFUNCTION CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE RISK DOCUMENTATION, POSSIBLE CAUSES OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MULTIPLE FACTORS, INCLUDING BUT NOT LIMITED TO, HIGHER WALL THICKNESS OF THE LEFT VENTRICLE AND/OR SURGICAL TECHNIQUE DURING IMPLANT. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED, DURING IMPLANT, THAT THE VENTRICULAR ASSIST DEVICE (VAD) APEX CORING TOOL WAS NOT CUTTING THE TISSUE EASILY. THE TOOL WAS RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 257171 | HEARTWARE VENTRICULAR ASSIST SYSTEM - APEX CORING TOOL | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1318 | 1377569 | 00888707000154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | MCS UNKNOWN VAD |