FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 7404312 · Received April 6, 2018

Report

Report Number
3005985723-2018-00220
Event Type
Malfunction
Date Received
April 6, 2018
Date of Event
March 15, 2018
Report Date
May 4, 2018
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K170581
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED ISSUE: THE HANDPIECE HAD A STUCK TRIGGER. PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE (B)(4) WERE MANUFACTURED UNDER LOT K06LC AND (B)(4) INCLUDING 4200386 WERE ACCEPTED INTO FINAL STOCK ON 2/2/16. A REVIEW OF (B)(4) REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N 209063, S/N (B)(6) IN PRODEX LOT K06LC SHOWS NO ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION. VISUAL INSPECTION: VISUAL INSPECTION SHOWS THE TRIGGER IS BINDING AND NOT ALLOWING THE TRIGGER TO FULLY RELEASE TO THE START POSITION. THE TRIGGER IS ABLE TO BE PULLED/FORCED TO THE START (FULLY RELEASED) POSITION. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED VISUAL AND FUNCTIONAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. MATERIAL ANALYSIS: MATERIAL ANALYSIS WAS NOT COMPLETED BECAUSE THE FAILURE WAS FUNCTIONAL. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION SHOWS THE HANDPIECE DOES NOT OPERATE WHEN THE TRIGGER IS IN THE BINDING STATE, SO THE COMPLAINT IS CONFIRMED. WHEN THE TRIGGER IS PULLED/ FORCED TO THE START (FULLY RELEASED) POSITION THE HANDPIECE STARTS AND ACCELERATES AS THE TRIGGER IS SQUEEZED. AFTER RELEASE OF THE TRIGGER, THE TRIGGER DOES NOT FULLY RELEASE. THE MOTOR CONTINUES TO SLOWLY TURN AS WOULD BE EXPECTED WITH A SLIGHTLY PRESSED TRIGGER. THE MOTOR SHUTS OFF WHEN THE TRIGGER IS PULLED/FORCED TO THE FULLY RELEASED POSITION. CONCLUSION: THE SYSTEM WILL NOT SPIN THE MOTOR IF THE HANDPIECE IS INITIALIZED WITH A PARTIALLY DEPRESSED TRIGGER. A BINDED TRIGGER LOOKS TO THE SYSTEM AS A PARTIALLY DEPRESSED TRIGGER. THE TRIGGER STICKS BECAUSE THE TRIGGER SHAFT IS MADE OF TWO COLLINEAR SHAFTS. THE JUNCTION OF THESE 2 SHAFTS SLIDES THROUGH A BORE. MISALIGNMENT OF THE 2 SHAFTS CAUSES THE TRIGGER TO BIND WHEN THE JUNCTION SLIDES THROUGH THE BORE.

Description of Event or Problem · 0

MICS TRIGGER MALFUNCTION. TKA CASE DELAYED 15 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE

Description of Event or Problem · 1

MICS TRIGGER MALFUNCTION. TKA CASE DELAYED 15 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245256 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 4200386 / 42020217

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization