FDA Adverse Event Malfunction Summary report: N

SMART MONITOR 2PS

MDR report key: 7402953 · Received April 6, 2018

Report

Report Number
3006182632-2018-00009
Event Type
Malfunction
Date Received
April 6, 2018
Date of Event
February 2, 2018
Report Date
March 28, 2018
Manufacturer
CIRCADIANCE, LLC
Product Code
NPF
UDI-DI
B0361030271
PMA / PMN Number
K061256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE NURSE CALL JACK IN THIS MONITOR WAS PHYSICALLY DAMAGED RENDERING THE OUTPUT INOPERABLE. CIRCADIANCE HAS RECENTLY RE-EVALUATED REPORTING DECISIONS RELATED TO PHYSICALLY DAMAGED NURSE CALL JACKS. CIRCADIANCE HAS MADE THE DECISION TO REPORT THESE INSTANCES OF DAMAGED NURSE CALL JACKS WHERE THE JACK IS STILL PHYSICALLY INTACT AND IT IS POSSIBLE THAT THE USER MAY HAVE PERCEIVED THAT THEY WERE STILL ABLE TO MAKE A CONNECTION TO THE JACK. THIS COMPLAINT WAS RECEIVED PRIOR TO THE RE-EVALUATION OF REPORTING, HOWEVER, THE EVENT OCCURRED RELATIVELY RECENTLY. THUS CIRCADIANCE HAS RE-CLASSIFIED THIS COMPLAINT AND HAS RETROSPECTIVELY FILED THIS REPORT TO BE CONSISTENT WITH THE RECENT INTERNAL CHANGES REGARDING REPORTING THESE EVENTS. THE PRIMARY ALARM IN THIS MONITOR AS WELL AS ALL OF THE MONITORING PARAMETERS WERE CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION.

Description of Event or Problem · 1

(B)(6) HOSPITAL SENT A SMART MONITOR 2PS MODEL NUMBER 1030271, SERIAL NUMBER (B)(4) IN FOR SERVICE REQUESTING THAT CIRCADIANCE CHECK THE OVERALL PERFORMANCE OF THE MONITOR. THE CUSTOMER DID NOT ALLEGE THAT THE PRODUCT WAS MALFUNCTIONING. THEY JUST REQUESTED THAT THE DEVICE BE EVALUATED. DURING SERVICE EVALUATION, THE SERVICE TECHNICIAN DETERMINED THAT THE NURSE CALL JACK WAS PARTIALLY DISLODGED FROM THE MAIN CIRCUIT BOARD RENDERING THE NURSE CALL OUTPUT NONFUNCTIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245008 SMART MONITOR 2PS INFANT APNEA MONITOR NPF CIRCADIANCE, LLC 1030271 B0361030271

Patients

Seq Age Sex Outcome Treatment
1 1 MO