FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 740287 · Received July 20, 2006

Report

Report Number
2243072-2006-00267
Event Type
Other
Date Received
July 20, 2006
Date of Event
June 19, 2006
Report Date
July 19, 2006
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT ERRATIC READINGS WITH HIS PLINK METER. ANALYSIS RUN ON CONSENSUS ERROR GRID (CEG) USING MEAN READINGS (326) AS REFERENCE POINT VS. BD READINGS (524, 128) YIELDED SOME DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 6064038

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other