FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0

MDR report key: 7401230 · Received April 5, 2018

Report

Report Number
3007042319-2018-01339
Event Type
Malfunction
Date Received
April 5, 2018
Date of Event
March 1, 2018
Report Date
August 15, 2019
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
UDI-DI
00888707000420
PMA / PMN Number
P100047
Removal / Correction Number
Z-1903-2018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTIONS: B3, B5, H6, H7, H9, H10. PRODUCT EVENT SUMMARY: THE CONTROLLER (B)(6) AND SIX (6) BATTERIES (B)(6) WERE RETURNED FOR EVALUATION. THE CONTROLLER AC ADAPTER (B)(6) WAS NOT RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED CONTROLLER AND BATTERIES REVEALED THAT THE DEVICES PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER, (B)(6), CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(6). DATA LOG FILES REVEALED A RELATIVE STATE OF CHARGE (RSOC) BETWEEN 101-201 INVOLVING (B)(6), WHICH IS INDICATIVE OF A COMMUNICATION ERROR. THE CONTROLLER LOG FILES ALSO REVEALED FIVE (5) CONTROLLER POWER-UPS WITH ASSOCIATED MOTOR STARTS LOGGED WITHIN THE ANALYZED PERIOD. THE POWER-UP EVENTS WERE LOGGED ON (B)(6) 2018 AT 13:10:00, (B)(6) 2018 AT 05:03:49, (B)(6) 2018 AT 06:28:11, (B)(6)2018 AT 07:51:16, AND (B)(6) 2018 AT 10:48:24. THE CONTROLLER WAS WITHOUT POWER FOR 22 SECONDS, 12 SECONDS, 10 SECONDS, 9 SECONDS, AND 11 SECONDS RESPECTIVELY. AS A RESULT, THE REPORTED EVENTS WERE CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE POWER SWITCHING EVENTS CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND BATTERIES. A POSSIBLE ROOT CAUSE OF THE LOSS OF POWER EVENTS CAN BE ATTRIBUTED TO DISCONNECTIONS OF BOTH POWER SOURCES AND/OR TO INTERMITTENT DISCONNECTIONS ON ONE OR BOTH POWER SOURCES. POSSIBLE ROOT CAUSES OF THE REPORTED COMMUNICATION ERROR CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS ON THE COMMUNICATION PINS OF THE CONTROLLER, THE CONTROLLER NOT RECEIVING RESPONSES FROM THE BATTERY, AND/OR DUE TO THE PACKET ERROR CHECKING METHOD DETECTING BIT ERRORS. AN INTERNAL INVESTIGATION WAS INITIATED TO CAPTURE EVENTS INVOLVING THE CONTROLLER LOSING POWER AND ANOTHER INTERNAL INVESTIGATION EXAMINED MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: D4: SERIAL#: (B)(6). H6: FDA CONCLUSION CODE(S): 4307 THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE CONTROLLER AND SIX BATTERIES WERE RETURNED FOR EVALUATION. THE CONTROLLER AC ADAPTER WAS NOT RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED CONTROLLER AND BATTERIES REVEALED THAT THE DEVICES PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER, (B)(4), CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15 MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(4).THE CONTROLLER LOG FILES ALSO REVEALED FIVE (5) CONTROLLER POWER-UPS WITH ASSOCIATED MOTOR STARTS LOGGED WITHIN THE ANALYZED PERIOD. THE CONTROLLER WAS WITHOUT POWER FOR 22 SECONDS, 12 SECONDS, 10 SECONDS, 9 SECONDS, AND 11 SECONDS RESPECTIVELY. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE POWER SWITCHING EVENTS CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND BATTERIES. A POSSIBLE ROOT CAUSE OF THE LOSS OF POWER EVENTS CAN BE ATTRIBUTED TO DISCONNECTIONS OF BOTH POWER SOURCES AND/OR TO INTERMITTENT DISCONNECTIONS ON ONE OR BOTH POWER SOURCES. AN INTERNAL INVESTIGATION WAS INITIATED TO CAPTURE EVENTS INVOLVING THE CONTROLLER LOSING POWER. MOMENTARY DISCONNECTIONS WERE INVESTIGATED INTERNALLY. (B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE RETURNED CONTROLLER PASSED VISUAL INSPECTION AND FUNCTIONAL TESTING. OTHER DEVICES INVOLVED IN THIS EVENT: BATTERY, BAT581485, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11 BATTERY, BAT582973, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11 BATTERY, BAT221536, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11 BATTERY, BAT221604, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11 BATTERY, BAT221538, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11 BATTERY, BAT222448, FDA RESULT CODE: 3233, FDA CONCLUSION CODE: 11. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CORRECTION: DATE OF EVENT. THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

LOG FILE REVIEW INDICATED ONE BATTERY ALSO HAD A COMMUNICATION ERROR.

Description of Event or Problem · 0

LOG FILE REVIEW INDICATED THAT THERE WAS A CONTROLLER POWER LOSS.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-06-30, UDI #: (B)(4), RETURN DATE: 2018-03-21, NO, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2017-06-30, (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-07-31, UDI #: (B)(4), RETURN DATE: 2018-03-21, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2017-07-31, (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-08-31, UDI #: (B)(4), RETURN DATE: 2018-08-21, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2017-08-31, (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-08-31, UDI #: (B)(4), RETURN DATE: 2018-08-21, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2017-08-31, (B)(4), HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2017-08-31, UDI #: (B)(4), RETURN DATE: 2018-08-21, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2016-08-31, (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY, BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2017-11-30, UDI #: (B)(4), RETURN DATE: 2018-08-21, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN, MFG DATE: 2016-11-30, (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ CAC ADAPTOR, CAC ADAPTOR/ (B)(4) / MODEL #: 1430DE, DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED FREQUENT BATTERY SWITCHING WHILE THE CONTROLLER WAS RUNNING. THE CONTROLLER, CONTROLLER AC ADAPTOR, AND SIX BATTERIES WILL BE EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243215 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0 VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1407DE 00888707000420

Patients

Seq Age Sex Outcome Treatment
1 (B)(6)