PERMOBIL C500
Report
- Report Number
- 1221084-2018-00025
- Event Type
- Injury
- Date Received
- April 5, 2018
- Date of Event
- March 20, 2018
- Report Date
- April 5, 2018
- Manufacturer
- PERMOBIL AB (PAB)
- Product Code
- ITI
- PMA / PMN Number
- K991658
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
THE COMPONENT FAILURE IS BEING ATTRIBUTED TO STRESSES BEING APPLIED OVER TIME LEADING TO EVENTUAL MATERIAL FATIGUE OF THE TOP PLATE. A CORRECTIVE AND PREVENTATIVE ACTION (CAPA (B)(4)) HAS BEEN IMPLEMENTED TO INVESTIGATE, CORRECT, AND MONITOR EFFECTIVENESS. AS A RESULT OF THE CAPA INVESTIGATION, IT WAS DECIDED THAT A THICKER 8 MM MATERIAL BE USED FOR THE TOP PLATE, REPLACING THE CURRENT 6 MM DESIGN. WITH THIS MATERIAL DESIGN CHANGE, IT WAS DETERMINED THE CHANGE TO THE THICKER MATERIAL IMPROVED THE OVERALL STRENGTH OF THE COMPONENT MAKING IT LESS SUSCEPTIBLE TO MATERIAL FATIGUE WHEN EXPOSED TO LOG TERM STRESSES. THE REDESIGNED COMPONENT HAS BEEN INSTALLED ON THE DEVICE, AND THE CHAIR RETURNED TO THE END-USER. THE DHR WAS REVIEWED AND CHAIR MET SPECIFICATION PRIOR TO DISTRIBUTION.
RECEIVED REPORT THAT WHILE END-USER WAS IN THEIR KITCHEN LOADING THE DISHWASHER, THE TOP PLATE OF THE ELEVATOR STARTED TO SEPARATE FROM THE INNER TUBE CAUSING THE SEATING TO LEAN FORWARD. THIS ALLEGEDLY CAUSED THE END-USER TO LOSE THEIR POSITIONING AND FALL OUT OF THE SEATING. IT WAS REPORTED THE END-USER SUSTAINED FRACTURED RIBS AS A RESULT OF THE FALL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243857 | PERMOBIL C500 | POWERED WHEELCHAIR | ITI | PERMOBIL AB (PAB) | C500 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |