FDA Adverse Event Injury Summary report: N

PERMOBIL C500

MDR report key: 7400730 · Received April 5, 2018

Report

Report Number
1221084-2018-00025
Event Type
Injury
Date Received
April 5, 2018
Date of Event
March 20, 2018
Report Date
April 5, 2018
Manufacturer
PERMOBIL AB (PAB)
Product Code
ITI
PMA / PMN Number
K991658
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
SERVICE PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE COMPONENT FAILURE IS BEING ATTRIBUTED TO STRESSES BEING APPLIED OVER TIME LEADING TO EVENTUAL MATERIAL FATIGUE OF THE TOP PLATE. A CORRECTIVE AND PREVENTATIVE ACTION (CAPA (B)(4)) HAS BEEN IMPLEMENTED TO INVESTIGATE, CORRECT, AND MONITOR EFFECTIVENESS. AS A RESULT OF THE CAPA INVESTIGATION, IT WAS DECIDED THAT A THICKER 8 MM MATERIAL BE USED FOR THE TOP PLATE, REPLACING THE CURRENT 6 MM DESIGN. WITH THIS MATERIAL DESIGN CHANGE, IT WAS DETERMINED THE CHANGE TO THE THICKER MATERIAL IMPROVED THE OVERALL STRENGTH OF THE COMPONENT MAKING IT LESS SUSCEPTIBLE TO MATERIAL FATIGUE WHEN EXPOSED TO LOG TERM STRESSES. THE REDESIGNED COMPONENT HAS BEEN INSTALLED ON THE DEVICE, AND THE CHAIR RETURNED TO THE END-USER. THE DHR WAS REVIEWED AND CHAIR MET SPECIFICATION PRIOR TO DISTRIBUTION.

Description of Event or Problem · 1

RECEIVED REPORT THAT WHILE END-USER WAS IN THEIR KITCHEN LOADING THE DISHWASHER, THE TOP PLATE OF THE ELEVATOR STARTED TO SEPARATE FROM THE INNER TUBE CAUSING THE SEATING TO LEAN FORWARD. THIS ALLEGEDLY CAUSED THE END-USER TO LOSE THEIR POSITIONING AND FALL OUT OF THE SEATING. IT WAS REPORTED THE END-USER SUSTAINED FRACTURED RIBS AS A RESULT OF THE FALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243857 PERMOBIL C500 POWERED WHEELCHAIR ITI PERMOBIL AB (PAB) C500 N/A

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization