FDA Adverse Event Injury Summary report: N

K2M MESA DEFORMITY

MDR report key: 7398769 · Received April 4, 2018

Report

Report Number
MW5076322
Event Type
Injury
Date Received
April 4, 2018
Date of Event
March 23, 2018
Report Date
April 3, 2018
Manufacturer
K2M
Product Code
OSH
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

PATIENT UNDERWENT A T10 TO S2A1 INSTRUMENTATION WITH K2M MESA INSTRUMENTATION. AT SURGERY ALL SCREWS WERE FINAL TIGHTENED ACCORDING TO MANUFACTURER SPECIFICS. AT 6 MONTHS, THE RODS HAVE LOST CONNECTION TO THE S2A1 SCREWS AT THE DISTAL END OF THE CONSTRUCT. THERE HAS BEEN NO ROD FAILURE. ALL HARDWARE WAS IN PERFECT POSITION. AT FOLLOW UP INTERVALS THE RODS REMAINED CONNECTED UNTIL 3 MONTHS AFTER SURGERY. THIS PATIENT HAD GOOD BONE QUALITY AND UNDERWENT A SCOLIOSIS CORRECTION WITH INTERNAL ROD REDUCTION TECHNIQUES. EOC SCREWS WERE USED AT S1 AND T10. RODS WERE FINAL TIGHTENED AT ALL LEVELS ACCORDING TO K2M SPECIFICS. THE LOCKING MECHANISM FOR THE K2M MESA SCREW MAY BE DEFICIENT. I HAVE ANOTHER CASE WITH THE EXACT SAME PROBLEM I WOULD BE HAPPY TO SHARE. IT IS MY INTENT THAT THE BIOMECHANICS OF THE FINAL TIGHTENING MECHANISM BE EVALUATED ON LONG CONSTRUCT FUSIONS IN THE LUMBAR SPINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241186 K2M MESA DEFORMITY LUMBAR PEDICLE SCREWS OSH K2M

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other| R