MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET
Report
- Report Number
- 1820334-2018-00905
- Event Type
- Malfunction
- Date Received
- April 4, 2018
- Date of Event
- March 23, 2018
- Report Date
- July 12, 2018
- Manufacturer
- COOK INC
- Product Code
- DYB
- UDI-DI
- 00827002355507
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION ¿ EVALUATION: A REVIEW OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD, DOCUMENTATION, QUALITY CONTROL, SPECIFICATIONS, AND A VISUAL INSPECTION OF THE RETURNED DEVICE WERE CONDUCTED DURING THE INVESTIGATION. THE VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THAT THE TIP OF THE INNER DILATOR WAS SPLIT AND FRAYED AS REPORTED. ADDITIONALLY, A DOCUMENT BASED INVESTIGATION EVALUATION WAS PERFORMED. THERE IS NO EVIDENCE TO SUGGEST THE PRODUCT WAS NOT MADE TO SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD SHOWED ONE NONCONFORMING EVENT WHICH COULD CONTRIBUTE TO THIS FAILURE MODE; HOWEVER, THE DEVICE WAS SCRAPPED AND NOT REPLACED. IT SHOULD BE NOTED THERE WERE NO OTHER REPORTED COMPLAINTS FOR THIS LOT NUMBER. BASED ON THE INFORMATION PROVIDED, THE EXAMINATION OF THE RETURNED PRODUCT, AND THE RESULTS OF OUR INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. WE WILL CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS. PER THE QUALITY ENGINEERING RISK ASSESSMENT, NO FURTHER ACTION IS REQUIRED. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, OR UNAVAILABLE. (B)(4). PMA/510(K) NUMBER = K133114. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.
IT WAS REPORTED THAT AFTER OPENING THE MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET, THE TIP WAS FRAYED (MEDWATCH 1820334-2018-00905). A SECOND MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET WAS THEN OPENED WITH THE SAME RESULTS MEDWATCH (1820334-2018-00904). NEITHER DEVICE WAS USED ON THE PATIENT. THE UNSPECIFIED PROCEDURE WAS SUCCESSFULLY COMPLETED USING A THIRD MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 238827 | MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET | DYB INTRODUCER, CATHETER | DYB | COOK INC | 00827002355507 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |