ADVIA CENTAUR XP TNI-ULTRA ASSAY
Report
- Report Number
- 1219913-2018-00124
- Event Type
- Malfunction
- Date Received
- April 3, 2018
- Date of Event
- March 16, 2018
- Report Date
- August 28, 2018
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MMI
- UDI-DI
- 00630414516219
- PMA / PMN Number
- K053020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SIEMENS FILED MDR 1219913-2018-00124 ON (B)(6)2018 FOR A DISCORDANT ADVIA CENTAUR XP CARDIAC TROPONIN I (TNI-ULTRA) PATIENT RESULT WHEN PERFORMING A CORRELATION RUN OF REAGENT LOT (010125) TO NEWER A REAGENT LOT (010127) RECEIVED. (B)(6)2018 - ADDITIONAL INFORMATION: MULTIPLE CUSTOMER COMPLAINTS REPORT A POSITIVE BIAS OF APPROXIMATELY 30% BETWEEN ADVIA CENTAUR TNI ULTRA REAGENT LOTS 010127 AND 010125 OR 010124 WHEN TESTING PATIENT SAMPLES AND QUALITY CONTROLS (QC) WITH CONCENTRATIONS APPROXIMATELY 0.10 NG/ML AND LOWER ON THE ADVIA CENTAUR XP/XPT. IN ADDITION, DEPRESSED QUALITY CONTROL RESULTS HAVE BEEN REPORTED ON BIO-RAD LEVEL 3 CONTROLS, WITH A NEGATIVE BIAS OF APPROXIMATELY -15%. IN SOME INSTANCES, CUSTOMER QC RESULTS FELL OUTSIDE THE CONTROL INSERT RANGES. SIMILAR REPORTS OF A LESSER MAGNITUDE HAVE ALSO BEEN REPORTED FOR TNI ULTRA REAGENT LOT 010127 ON ADVIA CENTAUR CP INSTRUMENTS. SIEMENS' COMPLAINT INVESTIGATION CONFIRMED CUSTOMER OBSERVATIONS OF BIAS WITH LOT 010127. INVESTIGATION DATA DEMONSTRATES A MEAN BIAS OF 33% BETWEEN TNI ULTRA REAGENT LOT 010127 AND 010125 FOR PATIENT SAMPLES IN THE RANGE OF 0.020 NG/ML AND 0.100 NG/ML ON THE ADVIA CENTAUR XP/XPT PLATFORM. ON THE ADVIA CENTAUR CP PLATFORM A MEAN BIAS OF 14% BETWEEN TNI ULTRA REAGENT LOT 010127 AND 010125 WAS OBSERVED FOR SAMPLES IN THE RANGE OF 0.020 NG/ML AND 0.100 NG/ML. THE ATELLICA IMMUNOASSAY ANALYZER TNI ULTRA ASSAY IS NOT IMPACTED, AS THE AFFECTED REAGENT LOT WAS NOT AVAILABLE FOR USE ON THIS PLATFORM. BIASES IN THE HIGH END OF THE ASSAY RANGE (CONCENTRATIONS APPROXIMATELY >20 NG/ML) WERE NEGATIVE, WHICH ALIGNS WITH CUSTOMER REPORTS OF A DROP IN CONTROL RECOVERY BEGINNING WITH REAGENT LOT 010127. SIEMENS TESTED QC MATERIAL AND MCMS, WHERE MEAN RESULTS MET ACCEPTABLE RANGES FOR EACH SAMPLE ON EVERY LOT AND PLATFORM BUT SHOWED SIMILAR BIASES TO PATIENT SAMPLES IN THE SAME AREA OF THE CURVE. FOR VERIFICATION PURPOSES, ADVIA CENTAUR TNI ULTRA LOTS 010125, 010128, 010130 AND 010132 WERE TESTED ON THE ADVIA CENTAUR XP AND ADVIA CENTAUR CP. PATIENT BIASES OBSERVED BETWEEN THESE LOTS WERE WITHIN EXPECTED PERFORMANCE BASED ON HISTORICAL PERFORMANCE OF THE ASSAY. ATELLICA IMMUNOASSAY ANALYZER REAGENT LOTS 010126, 010129 AND 010131 WERE ALSO TESTED AND BIASES OBSERVED BETWEEN THESE LOTS WERE SIMILAR TO OR BETTER THAN ADVIA CENTAUR SYSTEMS LOTS. IN ADDITION, A NORMAL RANGE STUDY CONFIRMED THAT ALL LOTS TESTED ARE MEETING THE REFERENCE INTERVAL AS STATED IN THE INSTRUCTIONS FOR USE AND DESIGN REQUIREMENTS (DIR). SIEMENS' INVESTIGATION CONCLUDED THAT THE BIAS OBSERVED WITH LOT 010127 IS A LOT SPECIFIC ISOLATED EVENT. SIEMENS' COMPLAINT INVESTIGATION IS COMPLETE. BASED ON THE INVESTIGATION RESULTS, SIEMENS CONCLUDES THAT ALTHOUGH ADVIA CENTAUR SYSTEMS TNI ULTRA REAGENT LOT 010127 HAS DEMONSTRATED A GREATER BIAS COMPARED TO OTHER LOTS, IT IS PERFORMING AS DESIGNED. THE INTENDED USE OF THE ASSAY IS NOT IMPACTED. SIEMENS HAS VERIFIED THAT THE PRODUCT IS MEETING ALL SPECIFICATIONS AND CLAIMS. THE AFFECTED REAGENT LOT 010127 EXPIRED ON 2018-07-08. MDR 1219913-2018-00125 WAS FILED FOR A DISCORDANT CORRELATION RESULT WITH REAGENT LOT 010128, AND MDR 1219913-2018-00125 SUPPLEMENTAL REPORT 1 WAS FILED FOR ADDITIONAL INFORMATION.
THE DISCORDANT ADVIA CENTAUR XP CARDIAC TROPONIN I (TNI-ULTRA) CORRELATION RESULT OBSERVED BY THE CUSTOMER IS BEING INVESTIGATED BY SIEMENS. THE SUMMARY AND EXPLANATION OF THE TEST SECTION OF THE INSTRUCTIONS FOR USE STATES THE FOLLOWING: "ALWAYS ANALYZE CTNI RESULTS IN THE CONTEXT OF TIME ELAPSED SINCE PATIENT PRESENTATION TO THE HOSPITAL. SERIAL SAMPLING IS RECOMMENDED TO DETECT THE TEMPORAL RISE AND FALL OF TROPONIN LEVELS CHARACTERISTIC OF MI. IN ACCORD WITH PUBLISHED RECOMMENDATIONS, SERIAL TESTING OF CTNI AT INTERVALS OF 2 TO 4 HOURS FOR UP TO 12 TO 24 HOURS IS SUGGESTED TO CORROBORATE A SINGLE CTNI RESULT. AN ELEVATED TROPONIN ALONE IS NOT SUFFICIENT TO MAKE THE DIAGNOSIS OF MI. OTHER MARKERS, SUCH AS CK-MB AND MYOGLOBIN, CAN BE USED IN CONJUNCTION WITH CTNI RESULTS IN AIDING THE DIAGNOSIS OF MI." THE INTERPRETATION OR RESULTS SECTION OF THE INSTRUCTION FOR USE STATES THE FOLLOWING: "RESULTS OF THIS ASSAY SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH PATIENT'S MEDICAL HISTORY, CLINICAL PRESENTATION AND OTHER FINDINGS." MDR 1219913-2018-00125 WAS FILED FOR A DISCORDANT CORRELATION RESULT WITH REAGENT LOT 010128.
THE CUSTOMER OBSERVED DISCORDANT ADVIA CENTAUR XP CARDIAC TROPONIN I (TNI-ULTRA) PATIENT RESULTS WHEN PERFORMING A CORRELATION RUN OF REAGENT LOT (010125) TO NEWER REAGENT LOTS (010127) AND (010128) RECEIVED. THE CUSTOMER HAS DECIDED NOT TO IMPLEMENT REAGENT (010127) FOR GENERAL USE, HOWEVER REAGENT LOT 010128 MEETS THE CUSTOMER'S QUALITY CONTROL (QC) RANGES AND IS NOW IN USE. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ADVIA CENTAUR XP TNI-ULTRA CORRELATION RESULT WITH REAGENT LOT 010127. MDR 1219913-2018-00125 WAS FILED FOR A DISCORDANT CORRELATION RESULT WITH REAGENT LOT 010128.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 237354 | ADVIA CENTAUR XP TNI-ULTRA ASSAY | TROPONIN IMMUNOASSAY | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | N/A | 010127 | 00630414516219 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |