G7 LINER
Report
- Report Number
- 0001825034-2018-02275
- Event Type
- Malfunction
- Date Received
- April 3, 2018
- Date of Event
- March 1, 2018
- Report Date
- June 26, 2018
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- PBI
- PMA / PMN Number
- PK121874
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT PRODUCTS: ITEM #: 010000997, ITEM NAME: G7 SCREW, LOT #: 3920093; ITEM #: 650-1057, ITEM NAME: BIOLOX DELTA HEAD, LOT #: 2904536; ITEM #: 650-1064, ITEM NAME: CER SLEEVE, LOT #: 2912968; ITEM #: 51-101070, ITEM NAME: TAPERLOC STEM, LOT #: 3963450; ITEM #: 010000664, ITEM NAME: G7 SHELL, LOT #: 6186608.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL INSPECTION OF THE RETURNED LINER SHOWS DEFORMATION TO THE SCALLOPS AROUND THE LINER, ALSO THE GROOVE FEATURE SHOWS INDENTATION ON THE OUTSIDE RADIUS OF THE LINER. NO FURTHER ANALYSIS WAS DONE AS THE LINER WAS IMPACTED. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED DURING A HIP ARTHROPLASTY THE LINER WOULD NOT SEAT WITH THE CUP, ANOTHER LINER WAS USED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 237155 | G7 LINER | PROSTHESIS, HIP | PBI | ZIMMER BIOMET, INC. | N/A | 6202181 |