FDA Adverse Event Malfunction Summary report: N

G7 LINER

MDR report key: 7393890 · Received April 3, 2018

Report

Report Number
0001825034-2018-02275
Event Type
Malfunction
Date Received
April 3, 2018
Date of Event
March 1, 2018
Report Date
June 26, 2018
Manufacturer
ZIMMER BIOMET, INC.
Product Code
PBI
PMA / PMN Number
PK121874
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT PRODUCTS: ITEM #: 010000997, ITEM NAME: G7 SCREW, LOT #: 3920093; ITEM #: 650-1057, ITEM NAME: BIOLOX DELTA HEAD, LOT #: 2904536; ITEM #: 650-1064, ITEM NAME: CER SLEEVE, LOT #: 2912968; ITEM #: 51-101070, ITEM NAME: TAPERLOC STEM, LOT #: 3963450; ITEM #: 010000664, ITEM NAME: G7 SHELL, LOT #: 6186608.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL INSPECTION OF THE RETURNED LINER SHOWS DEFORMATION TO THE SCALLOPS AROUND THE LINER, ALSO THE GROOVE FEATURE SHOWS INDENTATION ON THE OUTSIDE RADIUS OF THE LINER. NO FURTHER ANALYSIS WAS DONE AS THE LINER WAS IMPACTED. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED DURING A HIP ARTHROPLASTY THE LINER WOULD NOT SEAT WITH THE CUP, ANOTHER LINER WAS USED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.