FDA Adverse Event Injury Summary report: N

MENTOR SILTEX ROUND MODERATE PROFILE

MDR report key: 7392241 · Received April 3, 2018

Report

Report Number
1645337-2018-01909
Event Type
Injury
Date Received
April 3, 2018
Date of Event
March 1, 2018
Report Date
March 6, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001829
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 5/31/2018, THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE. WHITE MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING EXAMINATION, PRODUCT EVALUATION OBSERVED A RENT MEASURING APPROXIMATELY 0.4 CM NEXT TO AN AREA OF SILTEX CRACKING ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE FOUND. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ORIGIN OF WHITE MATERIAL OBSERVED IN THE DEVICE. THE DEVICE HISTORY RECORD ( DHR) OF LOT NUMBER 5874988 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES CAUSED BY ACUTE FOLDS (I.E., SEVERE ¿V¿ OR TIGHT FOLDS) IN THE RTV SHELL OF SILTEX BREAST IMPLANTS. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 425CC SALINE, CATALOG #3542670, LOT #5874988. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR OLD CAUCASIAN FEMALE PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH MENTOR SILTEX ROUND MODERATE PROFILE 425CC SALINE BREAST PROSTHESES. RIGHT BREAST PROSTHESIS DEFLATION WAS DETERMINED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236315 MENTOR SILTEX ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5874988 00081317001829

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention