PR4100 MICROPLATE READER US
Report
- Report Number
- 3022521-2018-00001
- Event Type
- Malfunction
- Date Received
- March 30, 2018
- Date of Event
- February 17, 2018
- Report Date
- December 21, 2018
- Manufacturer
- BIO-RAD LABORATORIES
- Product Code
- JJQ
- Removal / Correction Number
- 3022521-4/2/2018-002C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON (B)(6) 2018 BIO-RAD LABORATORIES TECHNICAL SUPPORT RECEIVED A CALL FROM A CUSTOMER STATING THAT THE D-LINK DUB-H4 USB HUB, USED WITH THE PR4100 MICROPLATE READER SYSTEM, STARTED SMOKING WHEN THE USER TURNED ON THE PR4100 COMPUTER. THE PR4100 MICROPLATE READER IS A 96-WELL ABSORBANCE READER FOR THE MEASUREMENT OF LIGHT ABSORBANCE (OPTICAL DENSITY-OD) OF LIQUID MEDIA. THE INSTRUMENT IS INTENDED TO BE USED PRIMARILY FOR IN VITRO DIAGNOSTIC ANALYSIS OF SAMPLES FROM THE HUMAN BODY TO OBTAIN INFORMATION ON PHYSIOLOGICAL AND PATHOLOGICAL STATES. THE PR4100 SYSTEM WAS TURNED OFF AND NO INJURY OCCURRED. BIO-RAD LABORATORIES TECHNICAL SUPPORT SENT THE CUSTOMER A TRIPP LITE USB HUB AS A REPLACEMENT. A FIELD CORRECTIVE ACTION HAS BEEN GENERATED TO REPLACE ALL EXISTING D-LINK HUBS THAT ARE CURRENTLY IN USE FOR PR4100 MICROPLATE READER CUSTOMERS. BIO-RAD LABORATORIES WILL SUBMIT A FOLLOW-UP REPORT UPON COMPLETION OF THE FIELD CORRECTIVE ACTION.
AS NOTED IN OUR INITIAL REPORT BIO-RAD LABORATORIES COMMITTED TO PROVIDING A FOLLOW-UP REPORT TO THE FDA. THERE HAS BEEN NO NEW OR ADDITIONAL INFORMATION, AND BIO-RAD LABORATORIES HAS COMPLETED THE FIELD ACTION FOR THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 229360 | PR4100 MICROPLATE READER US | SPECTROPHOTOMETER FOR CLINICAL USE | JJQ | BIO-RAD LABORATORIES |