FDA Adverse Event Malfunction Summary report: N

PR4100 MICROPLATE READER US

MDR report key: 7386745 · Received March 30, 2018

Report

Report Number
3022521-2018-00001
Event Type
Malfunction
Date Received
March 30, 2018
Date of Event
February 17, 2018
Report Date
December 21, 2018
Manufacturer
BIO-RAD LABORATORIES
Product Code
JJQ
Removal / Correction Number
3022521-4/2/2018-002C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

ON (B)(6) 2018 BIO-RAD LABORATORIES TECHNICAL SUPPORT RECEIVED A CALL FROM A CUSTOMER STATING THAT THE D-LINK DUB-H4 USB HUB, USED WITH THE PR4100 MICROPLATE READER SYSTEM, STARTED SMOKING WHEN THE USER TURNED ON THE PR4100 COMPUTER. THE PR4100 MICROPLATE READER IS A 96-WELL ABSORBANCE READER FOR THE MEASUREMENT OF LIGHT ABSORBANCE (OPTICAL DENSITY-OD) OF LIQUID MEDIA. THE INSTRUMENT IS INTENDED TO BE USED PRIMARILY FOR IN VITRO DIAGNOSTIC ANALYSIS OF SAMPLES FROM THE HUMAN BODY TO OBTAIN INFORMATION ON PHYSIOLOGICAL AND PATHOLOGICAL STATES. THE PR4100 SYSTEM WAS TURNED OFF AND NO INJURY OCCURRED. BIO-RAD LABORATORIES TECHNICAL SUPPORT SENT THE CUSTOMER A TRIPP LITE USB HUB AS A REPLACEMENT. A FIELD CORRECTIVE ACTION HAS BEEN GENERATED TO REPLACE ALL EXISTING D-LINK HUBS THAT ARE CURRENTLY IN USE FOR PR4100 MICROPLATE READER CUSTOMERS. BIO-RAD LABORATORIES WILL SUBMIT A FOLLOW-UP REPORT UPON COMPLETION OF THE FIELD CORRECTIVE ACTION.

Description of Event or Problem · 1

AS NOTED IN OUR INITIAL REPORT BIO-RAD LABORATORIES COMMITTED TO PROVIDING A FOLLOW-UP REPORT TO THE FDA. THERE HAS BEEN NO NEW OR ADDITIONAL INFORMATION, AND BIO-RAD LABORATORIES HAS COMPLETED THE FIELD ACTION FOR THIS INCIDENT.