DRIED POS MIC PANEL TYPE 20A
Report
- Report Number
- 2919016-2006-00009
- Event Type
- Malfunction
- Date Received
- July 6, 2006
- Date of Event
- June 15, 2006
- Report Date
- June 15, 2006
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- LTT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CONTACTED CUSTOMER TO OBTAIN CLINICAL ISOLATE FOR EVALUATION. ROUTINE MONITORING OF COMPLAIN HISTORY, AND PERFOMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. CLINICAL ISOLATE HAS NOT BEEN RECEIVED FROM THE CUSTOMER. INHOUSE TESTING OF THE ISOLATE IS ANTICIPATED, BUT NOT YET PERFORMED. OVERALL OXACILLIN PERFORMANCE OF DRIED POS PANELS IS ACCEPTABLE.
SINGLE ACCOUNT REPORTED TO DADE BEHRING THAT THEY OSBERVED A CLINICAL S.AUREUS ISOLATE OXACILLIN MIC DISCREPANCY. THE ACCOUNT OBTAINED OXACILLIN-SUSCEPTIBLE RESULTS ON THE DRIED POS MIC PANEL TYPE 20A PANEL AND OXACILLILN-RESISTANCT RESULTS ON A SECONDARY METHOD (CEFOXITIN DOSK SCREEN) ALSO PERFORMED FOR THE CLINICAL ISOLATE. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES TO THE PATIENT AS A RESULT OF THE FALSE SUSCEPTIBLE RESULTS BEING OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DRIED POS MIC PANEL TYPE 20A | ANTIMICROBIC SUSCEPT. TEST PANELS: GRAM POS. | LTT | DADE BEHRING, INC. | NA | 20007-04-06 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | UNK. | ||
| 2 | NA. |