FDA Adverse Event Malfunction Summary report: N

DRIED POS MIC PANEL TYPE 20A

MDR report key: 738069 · Received July 6, 2006

Report

Report Number
2919016-2006-00009
Event Type
Malfunction
Date Received
July 6, 2006
Date of Event
June 15, 2006
Report Date
June 15, 2006
Manufacturer
DADE BEHRING, INC.
Product Code
LTT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONTACTED CUSTOMER TO OBTAIN CLINICAL ISOLATE FOR EVALUATION. ROUTINE MONITORING OF COMPLAIN HISTORY, AND PERFOMANCE TRENDS TO ENSURE PERFORMANCE IS WITHIN CLAIMS. CLINICAL ISOLATE HAS NOT BEEN RECEIVED FROM THE CUSTOMER. INHOUSE TESTING OF THE ISOLATE IS ANTICIPATED, BUT NOT YET PERFORMED. OVERALL OXACILLIN PERFORMANCE OF DRIED POS PANELS IS ACCEPTABLE.

Description of Event or Problem · 1

SINGLE ACCOUNT REPORTED TO DADE BEHRING THAT THEY OSBERVED A CLINICAL S.AUREUS ISOLATE OXACILLIN MIC DISCREPANCY. THE ACCOUNT OBTAINED OXACILLIN-SUSCEPTIBLE RESULTS ON THE DRIED POS MIC PANEL TYPE 20A PANEL AND OXACILLILN-RESISTANCT RESULTS ON A SECONDARY METHOD (CEFOXITIN DOSK SCREEN) ALSO PERFORMED FOR THE CLINICAL ISOLATE. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES TO THE PATIENT AS A RESULT OF THE FALSE SUSCEPTIBLE RESULTS BEING OBTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRIED POS MIC PANEL TYPE 20A ANTIMICROBIC SUSCEPT. TEST PANELS: GRAM POS. LTT DADE BEHRING, INC. NA 20007-04-06

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN UNK.
2 NA.