MITROFLOW LXA
Report
- Report Number
- 3004478276-2018-00158
- Event Type
- Injury
- Date Received
- March 28, 2018
- Date of Event
- March 1, 2018
- Report Date
- April 18, 2018
- Manufacturer
- LIVANOVA CANADA CORP
- Product Code
- LWR
- UDI-DI
- 00896208000115
- PMA / PMN Number
- P060038
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
THE MANUFACTURING AND MATERIAL RECORDS FOR THE MITROFLOW BIOPROSTHETIC PERICARDIAL HEART VALVE, MODEL # LXA23, S/N (B)(4) , AS THEY PERTAIN TO THE REPORTED EVENT, WERE RETRIEVED AND REVIEWED BY QUALITY ENGINEERING AT LIVANOVA (B)(4) CORP. THE RESULTS CONFIRMED THAT THIS VALVE SATISFIED ALL MATERIAL, VISUAL, AND PERFORMANCE STANDARDS REQUIRED FOR A LXA23 MITROFLOW AORTIC PERICARDIAL HEART VALVE AT THE TIME OF MANUFACTURE AND RELEASE.
GROSS EXAMINATION OF THE RETURNED PROSTHESIS REVEALED DEFORMATION OF THE PROSTHESIS DUE TO INTRINSIC AND VEGETATING CALCIFICATION IN ALL LEAFLETS, AND A TEAR ON ONE OF THE LEAFLETS. PANNUS WAS PRESENT ON THE INFLOW SEWING RING OF THE VALVE. X-RAY ANALYSIS SHOWED LARGE, INTRINSIC CALCIFICATIONS IN ALL LEAFLETS. VISUAL EXAMINATION CONFIRMED THE PRESENCE OF A TEAR NEAR ONE OF THE POSTS. THE LEAFLETS WERE THICKENED DUE TO INTRINSIC AND VEGETATING CALCIFICATION, AND FOLDS WERE VISIBLE DUE TO IMPROPER LEAFLET MOTION DUE TO CALCIUM DEPOSITS. HISTOLOGICAL ANALYSIS ALSO SHOWED THICKENED LEAFLETS DUE TO INTRINSIC AND VEGETATING CALCIFICATION. HOMOGENATION OF THE COLLAGEN BUNDLES WAS EVIDENT. INFLAMMATORY CELLS WERE PRESENT ON TWO OF THE LEAFLETS, AND FIBROUS PANNUS DEPOSITION WAS VISIBLE ON THE MESOTHELIAL SURFACE OF ONE LEAFLET. FOCAL GRAM + BACTERIAL WAS PRESENT. AORTIC INSUFFICIENCY LIKELY RESULTED FROM A LEAFLET TEAR AND INADEQUATE LEAFLET COAPTATION DUE TO CALCIFICATION OF THE LEAFLETS. THERE WAS NO EVIDENCE OF ENDOCARDITIS IN THE RETURNED VALVE. STRUCTURAL DYSFUNCTION IS THE MAJOR CAUSE OF FAILURE OF BIOPROSTHETIC HEART VALVES, AND THE PRINCIPAL UNDERLYING PATHOLOGICAL PROCESS IS CUSPAL CALCIFICATION. CALCIFICATION CAN ALSO CAUSE STENOSIS DUE TO CUSPAL STIFFENING. CALCIFIC DEPOSITS ARE USUALLY LOCALIZED TO CUSPAL TISSUE (INTRINSIC CALCIFICATION). THE PATIENT'S CLINICAL CONDITIONS AND RISK FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT; HOWEVER, AS LIMITED PATIENT HISTORY WAS PROVIDED THE ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED.
A MITROFLOW LXA 23 MM VALVE WAS EXPLANTED AND REPLACED WITH PERCEVAL 23 MM VALVE. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 220423 | MITROFLOW LXA | TISSUE HEART VALVES | LWR | LIVANOVA CANADA CORP | LXA23 | 00896208000115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |