HANDPIECE MICS
Report
- Report Number
- 3005985723-2018-00206
- Event Type
- Malfunction
- Date Received
- March 28, 2018
- Date of Event
- March 23, 2018
- Report Date
- September 7, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K142530
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
REPORTED ISSUE: MICS FELL OF DURING THE CASE. PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K0A37 AND 25 INCLUDING 4202728 WERE ACCEPTED INTO FINAL STOCK ON 8/24/17. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO PARTS IN LOT NUMBER K0A37, P/N 209063 LOT K0A37 SHOWS NO OTHER COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION PRODUCT INSPECTION: VISUAL INSPECTION: VISUAL INSPECTION REVEALED NO PHYSICAL DAMAGE. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED. THE REPORTED PROBLEM WAS A FUNCTIONAL ISSUE. MATERIAL ANALYSIS: MATERIAL ANALYSIS WAS NOT COMPLETED BECAUSE THE FAILURE WAS FUNCTIONAL. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION REVEALED NO PHYSICAL DAMAGE AND THE PRODUCT FUNCTIONED AS INTENDED. ALL BOLTS AND SCREWS ON THE MICS WERE OBSERVED TO BE INTACT WHEN THE PRODUCT WAS INSPECTED. CONCLUSION: THE ALLEGED FAILURE WAS NOT CONFIRMED. NC/CAPA REVIEW: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN AN NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. THIS IS NC 1429704 AND CAPA 1452931.
THE BOLT ON MICS HANDPIECE BECAME LOOSE AND HANDPIECE FELL OFF INTRA OP. TKA CASE DELAYED 2 MINUTES.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE BOLT ON MICS HANDPIECE BECAME LOOSE AND HANDPIECE FELL OFF INTRA OP. TKA CASE DELAYED 2 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 221963 | HANDPIECE MICS | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 4202728 / 42040817 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |