DIMENSION VISTA®
Report
- Report Number
- 2517506-2018-00196
- Event Type
- Malfunction
- Date Received
- March 28, 2018
- Date of Event
- September 27, 2017
- Report Date
- March 28, 2018
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- NDR
- UDI-DI
- 00842768030314
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 2517506-03/22/2018-002-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED THAT DIMENSION VISTA® MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) LOT 7BDA81 MAY HAVE AN INCOMPLETE SLIT ON THE SEPTUM IN THE CAP OF THE VIAL. A DILUENT CAP WITH A DEFECTIVE SEPTUM HAS THE POTENTIAL TO CAUSE ERRONEOUSLY ELEVATED RESULTS IF A SAMPLE IS AUTODILUTED ONBOARD THE DIMENSION VISTA SYSTEM, DUE TO A REDUCED VOLUME OF DILUENT BEING PIPETTED FROM THE VIAL. MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) IS USED FOR DILUTIONS OF SAMPLES EXCEEDING THE ANALYTICAL MEASUREMENT RANGE WITH THE DIMENSION VISTA® ESTRADIOL (E2) FLEX REAGENT CARTRIDGE. URGENT MEDICAL DEVICE RECALL, VC-18-04.A.US DATED MARCH 21, 2018 AND URGENT FIELD SAFETY NOTICE VC-18-04.A.OUS DATED MARCH 2018 WERE ISSUED TO CUSTOMERS WHO HAD BEEN SHIPPED DIMENSION VISTA® MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) LOT 7BDA81 TO INFORM THEM OF THE ISSUE.
A FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED ON A CAP SURVEY SAMPLE ON THE DIMENSION VISTA 500 INSTRUMENT. THE RESULT WAS SCORED AS UNACCEPTABLE BY THE CAP. THERE IS NO INDICATION THAT PATIENT SAMPLES HAVE BEEN AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223295 | DIMENSION VISTA® | DIMENSION VISTA® MULTI 1 SAMPLE DILUENT | NDR | SIEMENS HEALTHCARE DIAGNOSTICS INC. | 7BDA81 | 00842768030314 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |