T-FIX RCG STERILE PAC

MDR report key: 7376039 · Received March 27, 2018

Report

Report Number
1219602-2018-00378
Event Type
Malfunction
Date Received
March 27, 2018
Date of Event
March 1, 2018
Report Date
April 26, 2018
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HRX
UDI-DI
03596010446114
PMA / PMN Number
K925573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOREIGN ZIP CODE: (B)(6).

Additional Manufacturer Narrative · 1

ONE 014771 T-FIX RCG STERILE PAC 2MM STERILE T FIX RCG DRILL WITH HUB RETURNED. THE COMPLAINT STATES: ¿IT WAS REPORTED THAT THE T-FIX WAS READY TO BE USED, PLUGGED ON THE POWER TOOL. THE T-FIX BROKE AT FULL SPEED, ENTERING THE CORACOID PROCESS. THE TIP OF THE MALE PART BROKE, JUST UNDER THE CONNECTION OF MALE/FEMALE PARTS. BROKEN PIECE WAS REMOVED FROM THE PATIENT. PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE.¿ THERE ARE SCRATCHES AND DINGS ON THE BLADE FLUTES. THE DRILL SHAFT IS SLIGHTLY BOWED AND HAS SPIRAL WEAR BANDS. THERE ARE ALSO RANDOM SCRATCHES IN OTHER LOCATIONS. THE DRILL¿S TAIL PORTION HAS BEEN FRACTURED FROM OVER TORQUE AT THE LOCATION WHERE THE 2MM DRILL HUB IS PERMANENTLY ATTACHED. PER THE IFU: ¿AS WITH ANY SURGICAL INSTRUMENT, CAREFUL ATTENTION SHOULD BE EXERCISED TO ENSURE THAT EXCESSIVE FORCE IS NOT PLACED ON THE INSTRUMENT. EXCESSIVE FORCE CAN RESULT IN INSTRUMENT FAILURE.¿ NO ROOT CAUSE ASSOCIATED WITH THE MANUFACTURE OF THIS DEVICE WAS ESTABLISHED. NO FURTHER INVESTIGATION WARRANTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE T-FIX WAS READY TO BE USED, PLUGGED ON THE POWER TOOL. THE T-FIX BROKE AT FULL SPEED, ENTERING THE CORACOID PROCESS. THE TIP OF THE MALE PART BROKE, JUST UNDER THE CONNECTION OF MALE/FEMALE PARTS. BROKEN PIECE WAS REMOVED FROM THE PATIENT. PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE. NO PATIENT INJURY REPORTED.