FDA Adverse Event Other Summary report: N

KODAK DIRECTVIEW CR CASSETTE

MDR report key: 736446 · Received July 1, 2006

Report

Report Number
1317307-2006-00006
Event Type
Other
Date Received
July 1, 2006
Date of Event
January 18, 2006
Report Date
June 29, 2006
Manufacturer
EASTMAN KODAK COMPANY
Product Code
IXA
PMA / PMN Number
K925997
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

KODAK RECEIVED THE SCREEN AND CASSETTE FOR EVALUATION. OUR EVALUATION INCLUDED CREATING AND PRINTING FLAT FIELD IMAGES AND A PHYSICAL EXAMINATION OF THE SCREEN. THE FLAT FIELD IMAGES WERE PRINTED AT AN EXTREME CONTRAST IN ORDER TO EHANCE ANY IMAGE ARTIFACTS. THE FLAT FIELD IMAGE SHOWED ARTIFACTS THAT WERE CLASSIFIED AS FINGERPRINT ARTIFACTS. THESE WERE LOCATED AT THE EDGES AND CORNERS OF THE IMAGE. THE PHYSICAL EXAMINATION SHOWED AN AREA WITH A SMALL ABRASION AT THE BOTTOM OF THE SCREEN. IN THE FLAT FIELD IMAGE THIS LOOKED LIKE A FOREIGN BODY. THE FINGERPRINT ARTIFACTS HAVE BEEN SEEN ON OTHER CR IMAGES AND ARE ASSOCIATED WITH HANDLING OF THE SCREENS BY BARE HANDS. MANY LOTIONS AND WATERLESS HAND CLEANERS CONTAIN CHEMICALS THAT CAN DAMAGE PHOSPHOR SCREENS. THE AFFECTED CR SCREEN WAS REMOVED AND REPLACED WITH A NEW CR SCREEN.

Description of Event or Problem · 1

A RADIOGRAPH TAKEN FOR A TRAUMA CASE SHOWED A FOREIGN BODY. ADDITIONAL RADIOGRAPHS WERE TAKEN TO LOCATE THE FOREIGN BODY AND THESE SHOWED NO FOREIGN BODY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KODAK DIRECTVIEW CR CASSETTE CR RIGID CASSETTE IXA EASTMAN KODAK COMPANY NA 1814

Patients

Seq Age Sex Outcome Treatment
1 Other