FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 7364236 · Received March 22, 2018

Report

Report Number
3004753838-2018-29473
Event Type
Malfunction
Date Received
March 22, 2018
Date of Event
February 21, 2018
Report Date
February 23, 2018
Manufacturer
DEXCOM, INC.
Product Code
PQF
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). "DEXCOM WAS MADE AWARE ON 02/23/2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION...THE COMPLAINT CONFIRMATION OF THE RECEIVER CEASING TO FUNCTION COULD NOT BE DETERMINED."

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON 02/23/2018 THAT ON (B)(6) 2018 THERE WAS UNEXPECTED RECEIVER SHUT-DOWN...THE COMPLAINT CONFIRMATION OF UNEXPECTED RECEIVER SHUT-DOWN COULD NOT BE DETERMINED. THE COMPLAINT RECEIVER DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXTERNALLY VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. THE RECEIVER WAS ABLE TO BE CHARGED AND REBOOTED. FUNCTIONAL TESTING WAS PERFORMED AND IT PASSED. THE RECEIVER LOG WAS DOWNLOADED, THE DATA CONTAINS NO ISSUES RELATED TO COMPLAINT. THE RECEIVER CASE WAS OPENED AND THE INTERNAL INSPECTION PASSED. THE REPORTED EVENT OF UNEXPECTED RECEIVER SHUTDOWN WAS NOT CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON 02/23/2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER CEASING TO FUNCTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
206806 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQF DEXCOM, INC. MT22719-2 5230569

Patients

Seq Age Sex Outcome Treatment
1 9 YR