FDA Adverse Event Death Summary report: N

CS300

MDR report key: 7362897 · Received March 22, 2018

Report

Report Number
2249723-2018-00465
Event Type
Death
Date Received
March 22, 2018
Date of Event
February 17, 2018
Report Date
March 22, 2018
Manufacturer
DATASCOPE MAHWAH
Product Code
DSP
PMA / PMN Number
K063525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SY
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCTION DEVICE HISTORY RECORD (DHR) FOR THIS IABP UNIT CANNOT BE REVIEWED PER THE SOP SINCE THE SERIAL NUMBER FOR THE UNIT WAS NOT PROVIDED. (B)(6). (B)(4). THE CUSTOMER HAS NOT REPORTED AN ALLEGED MALFUNCTION OF THE IABP AND THEREFORE HAS NOT REQUESTED GETINGE TO EVALUATE THE UNIT. THE DETAILS OF THE INVOLVED IABP HAVE NOT BEEN PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CS300 INTRA-AORTIC BALLOON (IABP) THERAPY, SOME PROBLEMS OCCURRED DURING THE INSERTION OF TWO INTRA-AORTIC BALLOON (IAB) CATHETERS IN THE CARDIAC CENTER. THE FIRST IAB CATHETER HAD ITS AXES BROKEN AFTER COMPLETION OF THE INSERTION PROCEDURE, WHICH CAUSED INFUSION OF BLOOD INTO THE BALLOON. THE CUSTOMER HAS CONCERNS THAT THIS MAY HAVE CAUSED THE DEATH OF THE PATIENT. THE SECOND IAB CATHETER HAD ITS AXES BROKEN DURING THE INSERTION DUE TO THE SHEATH WHICH WAS TOO TIGHT TO INSERT THE CATHETER EASILY INTO IT. THE MEDICAL STAFF USED A THIRD BALLOON CATHETER WHICH WAS INSERTED CORRECTLY, AND THERAPY WAS CONTINUED. THE PATIENT DIED 8 HOURS LATER AND IT IS UNKNOWN WHETHER THE CAUSE OF DEATH WAS DUE TO THE CATHETER OR DUE TO THE CLINICAL STATUS OF THE PATIENT. THERE WAS NO ALLEGED MALFUNCTION OF THE IABP. THIS EVENT IS RELATED TO IAB REPORTS TO BE SUBMITTED UNDER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
205016 CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE MAHWAH N/A N/A

Patients

Seq Age Sex Outcome Treatment
1 60 YR Death LINEAR 7.5 FR. 40CC IAB