FDA Adverse Event Malfunction Summary report: N

DEFENSE BLOOD STANDARD SYSTEM

MDR report key: 736033 · Received July 20, 2005

Report

Report Number
1119523-2005-00003
Event Type
Malfunction
Date Received
July 20, 2005
Date of Event
December 15, 2004
Report Date
July 15, 2005
Manufacturer
DOD/DMLSS/PROJECT OFFICE
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ATTACHMENT #1: DTB 05-13. ATTACHMENT #2: DBO 2619. A SCRIPT IS BEING DEVELOPED TO ADDRESS SITUATIONS WHERE A FAILED UPLOAD HAS ALREADY OCCURRED AND HAS GONE UNDETECTED AND TO FIX POST-MERGE PERSON RECORDS THAT COULD POTENTIALL REPRODUCE THE UPLOAD PROBLEM. THIS SCRIPT WILL ALSO ELIMINATE THE NEED TO CONTINUE USING THE WORKAROUND PROVIDED IN THIS DTB. ONCE THE SCRIPT IS AVAILABLE THIS DTB WILL BE UPDATED NOTIFYING USERS. A TROUBLE TICKET WILL NEED TO BE OPENED WITH THE MHS HELP DESK FOR AN INVESTIGATION TO BE PERFORMED TO DETERMINE IF A DATAMOD IS REQUIRED TO CORRECT PROBLEMS THAT ARE ALREADY IN YOUR SYSTEM.

Description of Event or Problem · 1

UPON UPLOAD TO THE IN-HOUSE SERVER, DBSS FAILS TO ASSIGN A NEW MOBILE DONOR REGISTRATION, PHLEBOTOMY AND UNIT INFORMATION TO THE ACTIVE RECORD ON THE SERVER. THIS PROBLEM ONLY OCCURS IN CASES WHERE THE DONOR WAS PART OF A PERSON MERGE PRIOR TO THE MOBILE DOWNLOAD/UPLOAD PROCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEFENSE BLOOD STANDARD SYSTEM BLOOD ESTABLISHMENT COMPUTER SOFTWARE MMH DOD/DMLSS/PROJECT OFFICE * *

Patients

Seq Age Sex Outcome Treatment
1 * Other