FDA Adverse Event Malfunction Summary report: N

T:SLIM G5 SYSTEM

MDR report key: 7357447 · Received March 21, 2018

Report

Report Number
3013756811-2018-10113
Event Type
Malfunction
Date Received
March 21, 2018
Date of Event
March 4, 2018
Report Date
March 21, 2018
Manufacturer
TANDEM DIABETES CARE
Product Code
OYC
PMA / PMN Number
P140015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE T:SLIM X2 G5 PUMP USER GUIDE STATES: YOUR T:SLIM X2 PUMP CAN STOP INSULIN DELIVERY AND ALERT YOU (OR WHOEVER IS WITH YOU) IF THERE HAS BEEN NO INTERACTION WITH THE PUMP WITHIN A SPECIFIED PERIOD OF TIME. THE DEFAULT FOR THIS ALARM IS PRE-SET TO 12 HOURS. YOU CAN SET IT ANYWHERE BETWEEN 5 AND 24 HOURS, OR OFF. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER WAS INSTRUCTED TO PLUG THE PUMP INTO A POWER SOURCE FOR AT LEAST 30 MINUTES. CUSTOMER ACKNOWLEDGED. REPORTEDLY, THE ALERT HAD NOT REOCCURRED AFTER CHARGING THE PUMP. IN ADDITION, IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN AUTO-OFF ALARM. TANDEM TECHNICAL SUPPORT EDUCATED THE CUSTOMER ON THE AUTO-OFF SAFETY FEATURE. THE CUSTOMER¿S BLOOD GLUCOSE WAS 150-235 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200269 T:SLIM G5 SYSTEM CONTINUOUS GLUCOSE MONITOR OYC TANDEM DIABETES CARE 1000096

Patients

Seq Age Sex Outcome Treatment
1 60 YR