FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 7356555 · Received March 21, 2018

Report

Report Number
3008382007-2018-00894
Event Type
Malfunction
Date Received
March 21, 2018
Report Date
March 16, 2018
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
Z-1074-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER CAREGIVERS

Narratives

Description of Event or Problem · 1

ON (B)(6) 2018, THE REPORTER CONTACTED LIFESCAN (B)(6), ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿368, 212, 182, 176 AND 189 MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.