CATALYS SYSTEM
Report
- Report Number
- 3005675890-2018-00012
- Event Type
- Malfunction
- Date Received
- March 21, 2018
- Date of Event
- February 21, 2018
- Report Date
- April 23, 2018
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- OOE
- UDI-DI
- 05050474609792
- PMA / PMN Number
- K121091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
JOHNSON & JOHNSON SURGICAL VISION, INC. INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
DEVICE EVALUATION: A REVIEW OF THE RECORDS RELATED TO THE DEVICE THAT INCLUDED LABELING, MANUALS, TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THE TREND REVIEW SHOWS THAT THERE IS NOT SIGNIFICANT CHANGE OVER HISTORICAL COMPLAINT LIMITS AND NO RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. THERE WAS NO PRODUCT DEFICIENCY IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
ACCOUNT REPORTED THAT THERE WAS SUCTION LOSS WHILE LASER FIRED DURING AT THE END OF FRAGMENTATION. ACCOUNT WAS ABLE TO COMPLETE THE TREATMENT OF ARCUATE INCISIONS WITH A DIFFERENT LIQUID OPTICS INTERFACE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 201272 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC. | CATALYS-U | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |