FDA Adverse Event Malfunction Summary report: N

TOSOH HLC-723G8 ANALYZER

MDR report key: 7356159 · Received March 20, 2018

Report

Report Number
8031673-2018-03317
Event Type
Malfunction
Date Received
March 20, 2018
Date of Event
February 3, 2017
Report Date
March 20, 2018
Manufacturer
TOSOH CORPORATION
Product Code
LCP
PMA / PMN Number
K071132
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ON 03-FEB-2017, AN FSE CONDUCTED A FOLLOW-UP WITH THE CUSTOMER OVER-THE-PHONE TO ADDRESS THE REPORTED EVENT. THE FSE GUIDED THE CUSTOMER ON HOW TO REPLACE SAMPLE NEEDLE AND PRIMED SYSTEM. NO FURTHER ACTION WAS REQUIRED BY FIELD SERVICE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS THE SAMPLE NEEDLE. TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. THIS REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW CONDUCTED UNDER CAPA-2017-0007. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE IMPORTER OR MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 0

ON (B)(6) 2017, A CUSTOMER REPORTED 706 LARGE SYRINGE ERRORS WITH GRINDING SOUND WITH THEIR G8 ANALYZER. THE CUSTOMER WAS UNABLE TO RUN HBA1C PATIENT SAMPLES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAYED REPORTING OF HBA1C PATIENT RESULTS. THERE WAS NO INDICATION OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197802 TOSOH HLC-723G8 ANALYZER G8 LCP TOSOH CORPORATION G8

Patients

Seq Age Sex Outcome Treatment
1