AIA-900
Report
- Report Number
- 8031673-2018-02958
- Event Type
- Malfunction
- Date Received
- March 20, 2018
- Date of Event
- January 27, 2017
- Report Date
- March 20, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
H10 ADDITIONAL NARRATIVE: THIS REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW CONDUCTED UNDER CAPA-2017-0007. TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE IMPORTER OR MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT. H.3. DEVICE EVALUATION BY MANUFACTURER: THE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON 27 JANUARY 2017. THE FSE REPLACED THE WASH SUPPLY VALVES. THE ENGINEER REPORTED AFTER RUNNING QUALITY CONTROL THE RECOVERY PRECISION WAS GOOD.THE AIA-900 WAS WORKING AS INTENDED. NO FURTHER ACTION REQUIRED BY FSE. THE PROBABLE CAUSE OF THE EVENT WAS THE WASH SUPPLY VALVES.
ON (B)(6) 2017 A CUSTOMER REPORTED QUALITY CONTROL RESULTS FOR FSH DRIFTING LOW WITH THEIR AIA-900 ANALYZER. THE CUSTOMER WAS UNABLE TO RUN SAMPLES AND REQUESTED SERVICE TO ADDRESS THE REPORTED EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAY REPORTING FSH PATIENT RESULTS. THERE WAS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE EVENT CONSEQUENCE DUE TO THE DELAY OF REPORTING PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 197764 | AIA-900 | AIA-900, PRODUCT CODE: KHO | KHO | TOSOH CORPORATION | AIA-900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |