FDA Adverse Event Malfunction Summary report: N

SPYSCOPE DS ACCESS & DELIVERY CATHETER

MDR report key: 7352273 · Received March 20, 2018

Report

Report Number
3005099803-2018-00674
Event Type
Malfunction
Date Received
March 20, 2018
Date of Event
February 23, 2018
Report Date
February 23, 2018
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FBN
UDI-DI
08714729863236
PMA / PMN Number
K142922
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.

Additional Manufacturer Narrative · 1

A VISUAL ASSESSMENT WAS PERFORMED AFTER DISINFECTION. AS RECEIVED, THE WORKING CHANNEL SLEEVE PROTRUDED, CONFIRMING THE COMPLAINT. THE DISTAL TIP WAS CUT. THE DISTAL CAP WAS REMOVED. THE CATHETER WAS CUT OPEN USING THE CUTTING FIXTURE. THE WORKING CHANNEL SLEEVE WAS REMOVED. WITNESS MARKS WERE NOTED ON THE PEBAX. THE WHITE AREAS ALONG BOND A APPEAR TO SHOW EVIDENCE OF ADHESION. THE COMPLAINT WAS CONSISTENT WITH THE REPORTED EVENT OF WORKING CHANNEL SLEEVE PROTRUDING. BASED ON THE INVESTIGATION RESULTS, THE UNDERLYING CAUSE OF WORKING CHANNEL SLEEVE PROTRUSION IS AN INSUFFICIENT BOND, PARTICULARLY THE SECOND HEAT CYCLE OF THE WORKING CHANNEL SLEEVE BONDING PROCESS [BOND B]. THEREFORE, THE COMPLAINT INVESTIGATION CONCLUSION CODE SELECTED FOR THE WORKING CHANNEL SLEEVE PROTRUSION ISSUE IS MANUFACTURING PROCESS DESIGN. AN INVESTIGATION IS UNDERWAY TO ADDRESS THIS ISSUE. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING AN INDETERMINATE DIAGNOSIS OF STENOSIS BY BIOPSY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. THERE WERE NO REPORTED ISSUES WITH THE ACCESSORY DEVICE. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING AN INDETERMINATE DIAGNOSIS OF STENOSIS BY BIOPSY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. THERE WERE NO REPORTED ISSUES WITH THE ACCESSORY DEVICE. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
198065 SPYSCOPE DS ACCESS & DELIVERY CATHETER CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID FBN BOSTON SCIENTIFIC - SPENCER M00546600 21668796 08714729863236

Patients

Seq Age Sex Outcome Treatment
1