PRECISION SPECTRA
Report
- Report Number
- 3006630150-2018-01010
- Event Type
- Malfunction
- Date Received
- March 19, 2018
- Date of Event
- February 28, 2018
- Report Date
- March 19, 2018
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- UDI-DI
- 08714729821526
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-8216-50 SERIAL #: (B)(4) DESCRIPTION: ARTISAN SURGICAL LEAD, 50CM MODEL #: SC-8216-70 SERIAL #: (B)(4) DESCRIPTION: ARTISAN SURGICAL LEAD, 70CM THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE KEPT BY THE MEDICAL FACILITY. A REVIEW OF THE MANUFACTURING DOCUMENTATION FOR THE IPG AND LEADS REVEALED THAT NO ANOMALIES OR DEVIATIONS POTENTIALLY RELATED TO THE EVENT OCCURRED DURING MANUFACTURING.
A REPORT WAS RECEIVED THAT THE PHYSICIAN USED ELECTROCAUTERY AND FRIED THE LEAD DURING FUSION SURGERY. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE LEAD AND IPG WERE REPLACED DUE TO DIRECT ELECTROCAUTERY ON THE LEAD. NO DEVICE MALFUNCTION WAS SUSPECTED. ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND CURRENT COMPANY LABELING WARNS AGAINST THE USE OF ELECTROCAUTERY. (PHYSICIAN¿S IMPLANT MANUAL (B)(4)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193282 | PRECISION SPECTRA | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-1132 | NA | 08714729821526 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |