FDA Adverse Event Malfunction Summary report: N

DHS®/DCS® GUIDE SHAFT

MDR report key: 7351604 · Received March 19, 2018

Report

Report Number
2939274-2018-51110
Event Type
Malfunction
Date Received
March 19, 2018
Date of Event
February 19, 2018
Report Date
February 19, 2018
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
FZX
UDI-DI
10886982193247
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR THE SUBJECT DEVICE: PART NO.: 338.210, LOT NO.: 1203360, MANUFACTURING LOCATION: (B)(4), RELEASE TO WAREHOUSE DATE: 24.SEP.2003. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. A PRODUCT DEVELOPMENT INVESTIGATION WAS PERFORMED FOR THE SUBJECT DEVICE: DEVICE CONDITION: THE DHS/DCS GUIDE SHAFT, 338.21, WAS RECEIVED WITH A SLIGHT OFF AXIS BEND. THE ORIGIN OF THE BEND IS LOCATED APPROXIMATELY 25MM FROM THE DISTAL END OF THE DEVICE (CALIPERS: CA102P). THE DEVICE WAS FUNCTIONALLY TESTED WITH THE RETURNED MATING COUPLING SCREW. THE DEVICES WERE FOUND TO ASSEMBLE AND DISASSEMBLE BUT ADDITIONAL RESISTANCE WAS ENCOUNTERED DURING ASSEMBLED AND DISASSEMBLY DUE TO THE BENT CONDITION. THE PROXIMAL END OF THE SHAFT ALSO SHOWS SCRAPING CONSISTENT WITH HAVING BEEN GRIPPED WITH ANOTHER TOOL. THE BALANCE OF THE DEVICE SHOWS SURFACE WEAR CONSISTENT WITH USE AND WHICH WOULD NOT IMPACT THE FUNCTIONALITY. THUS, THE COMPLAINT CONDITION IS CONFIRMED AND CONSISTENT WITH THE REPORTED CONDITION. REPLICATION OF THE COMPLAINT CONDITION IS NOT APPLICABLE AS THE DEVICE IS ALREADY BENT. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE LOT WENT THROUGH THE REQUIRED STEPS DURING THE INSPECTION AT THE TIME OF MANUFACTURING AND THE DHR RECORDS SHOWED NO ISSUES CONCERNING THE MATERIAL OR MATERIAL CONDITIONING. RELEASE TO WAREHOUSE DATE: 24.SEP.2003. THE FOLLOWING DRAWINGS, REFLECTING THE CURRENT AND MANUFACTURED REVISION, WERE REVIEWED; DHS/DCS GUIDE SHAFT. THE OUTER DIAMETER OF THE SHAFT ON EACH SIDE OF THE BEND WAS INSPECTED. THE DEVICE DIAMETER MEASURED 7.04MM/7.00MM WHICH IS WITHIN THE SPECIFICATION OF 7.05MM +/-0.05 PER DRAWING (CALIPERS: CA102P). THE INNER DIAMETER OF THE SHAFT ON THE DISTAL END WAS FOUND TO MEASURE 5.13MM WHICH IS WITHIN THE SPECIFICATION OF 5.1MM +/-0.05 PER DRAWING (CALIPERS: CA102P). DURING THE INVESTIGATION NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. NO DEFINITIVE ROOT CAUSE WAS ABLE TO BE DETERMINED AS CIRCUMSTANCES SURROUNDING THE INTRAOPERATIVE EVENT AND THE SUBSEQUENT ISSUES DURING DISASSEMBLY ARE UNKNOWN. HOWEVER, THE RETURNED CONDITION IS CONSISTENT WITH AN APPLICATION OF EXCEEDING OFF-AXIS FORCE. THE COMPLAINT CONDITION IS CONFIRMED AS THE GUIDE SHAFT WAS RECEIVED WITH A SLIGHT OFF AXIS BEND. A DEVICE HISTORY RECORD (DHR) REVIEW, DEVICE INSPECTION, FUNCTIONAL TEST, AND DRAWING REVIEW WERE PERFORMED AS PART OF THIS INVESTIGATION. THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION AND NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. NO NEW MALFUNCTIONS WERE OBSERVED DURING THE COURSE OF THIS INVESTIGATION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. PATIENT WEIGHT IS NOT PROVIDED FOR REPORTING. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. (B)(4). DEVICE HISTORY RECORDS REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THE: DHR REVIEW DHR REVIEW, PART NO.: 338.210, LOT NO.: 1203360, PART # 338.21, LOT # 4683918, SERIAL/SUPPLIER LOT (B)(4), MANUFACTURING LOCATION: (B)(4), RELEASE TO WAREHOUSE DATE: 24.SEP.2003. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. NO NCRS WERE GENERATED DURING PRODUCTION.

Additional Manufacturer Narrative · 1

DATE RETURNED TO MANUFACTURER: SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER AN INITIAL SURGERY FOR IMPLANTATION OF A DYNAMIC HIP SYSTEM (DHS) INTO THE PROXIMAL FEMUR ON (B)(6) 2018, IN STERILE PROCESSING, THE CONNECTING SCREW BROKE INTO TWO (2) PIECES AND THE GUIDE SHAFT, STILL IN ONE PIECE, WAS BENT TRYING TO GET THEM APART. DURING THE SURGERY, AS THE SURGEON WAS REMOVING THE CONNECTING SCREW AND GUIDE SHAFT FROM THE LAG SCREW, THE TWO INSTRUMENTS BECAME BENT (IT WAS REPORTED THAT THE SURGEON DID NOT DISCONNECT THE INSTRUMENTS PRIOR TO TAKING THEM OUT) AND THE SURGEON HAD TO USE PLIERS TO TAKE THE INSTRUMENTS OUT. THIS COMPLAINT IS LINKED WITH (B)(4) FOR THE INTRAOPERATIVE EVENTS. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS 2 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191892 DHS®/DCS® GUIDE SHAFT GUIDE FZX WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4683918 10886982193247

Patients

Seq Age Sex Outcome Treatment
1 18 YR