TOSOH HLC-723G8 ANALYZER G8
Report
- Report Number
- 8031673-2018-03200
- Event Type
- Malfunction
- Date Received
- March 19, 2018
- Date of Event
- December 14, 2015
- Report Date
- March 19, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- LCP
- PMA / PMN Number
- K071132
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ON (B)(6) 2015, AN FSE CONDUCTED TROUBLESHOOTING OVER THE PHONE WITH THE CUSTOMER TO ADDRESS THE REPORTED EVENT. THE CUSTOMER WAS ASKED TO REPLACE THE COLUMN BUT THE TECH AT THE FACILITY WAS NOT FAMILIAR WITH THE PROCESS. THE CUSTOMER STATED THE COLUMN WILL BE CHANGED IN THE MORNING BY ANOTHER TECH. THE NEXT DAY, THE FSE CALLED BACK TO ASSIST THE CUSTOMER WITH FLUSHING OUT THE SYSTEM AND THE CUSTOMER REPORTED NORMAL RETENTION TIMES WITH NO ABNORMAL PEAKS. THE INSTRUMENT WAS VERIFIED AS OPERATIONAL. THE G8 ANALYZER WAS FUNCTIONING AS INTENDED AND THE ISSUE CLEARED. NO FURTHER ACTION REQUIRED BY FIELD SERVICE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS PEEK TUBING. TOSOH BIOSCIENCE, INC. IS SUBMITTNG ON BEHALF OF THE FOREIGN MANUFACTURER TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. THIS REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW CONDUCTED UNDER CAPA(B)(4).
ON (B)(6) 2015, A CUSTOMER REPORTED HIGHER THAN NORMAL RETENTION TIMES WITH THEIR G8 ANALYZER. THE CUSTOMER TRIED ADJUSTING THE FLOW RATE BUT DID NOT SEE ANY IMPROVEMENT IN THE RETENTION TIMES. THE CUSTOMER IS UNABLE TO RUN HBA1C PATIENT SAMPLES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAYED REPORTING OF HBA1C PATIENT RESULTS. THERE WAS NO INDICATION OF ANY PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 192474 | TOSOH HLC-723G8 ANALYZER G8 | TOSOH HLC-723G8 ANALYZER G8 | LCP | TOSOH CORPORATION | G8 |