FDA Adverse Event Injury Summary report: N

CIRRUS

MDR report key: 7348646 · Received March 19, 2018

Report

Report Number
7348646
Event Type
Injury
Date Received
March 19, 2018
Date of Event
March 4, 2018
Report Date
March 9, 2018
Manufacturer
TOLLOS, INC.
Product Code
FSA
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

WHILE TRANSFERRING A PATIENT FROM THE BED TO A STRETCHER, THE LIFT SNAPPED, CAME OFF THE TRACK IN THE CEILING AND FELL ONTO THE PATIENT. A TRANSPORTER WAS WITH THE PATIENT AND ATTEMPTED TO STOP IT FROM HITTING THE PATIENT. THE TRANSPORTER SUFFERED A BROKEN ARM AS A RESULT. AFTERWARDS THE PATIENT COMPLAINED OF HIP PAIN. X-RAYS WERE DONE TO CONFIRM NO FURTHER INJURY. BIOMED CONFIRMED THAT THE SAFETY PIN ON THE TRACK WAS MISSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192141 CIRRUS LIFT, PATIENT, NON-AC-POWERED FSA TOLLOS, INC. 600

Patients

Seq Age Sex Outcome Treatment
1 64 YR NOT APPLICABLE.