FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE

MDR report key: 7345834 · Received March 16, 2018

Report

Report Number
1213809-2018-00164
Event Type
Malfunction
Date Received
March 16, 2018
Date of Event
March 6, 2018
Report Date
April 5, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
00382903096466
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: THE INITIAL MDR REPORTED THE LOT NUMBER AS UNKNOWN. THE LOT # HAS BEEN IDENTIFIED AS 8011965. D.4. LOT NUMBER: 8011965. H.6. INVESTIGATION RESULTS: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. DHR REVIEW FOR BATCH 8011965 (P/N 309646): MANUFACTURING DATES: 01/22/2018 TO 01/23/2018. BATCH QUANTITY WAS (B)(4). PRINTING AND ASSEMBLY RECORDS WERE REVIEWED AS PART OF THIS DHR REVIEW. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 8011965 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. THE REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. NO CAPA IS REQUIRED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE PHLEBOTOMIST THAT THE RUBBER STOPPER OF THE BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE ARE POPPING OFF DURING USE. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187230 BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 8011965 00382903096466

Patients

Seq Age Sex Outcome Treatment
1 Other