BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE
Report
- Report Number
- 1213809-2018-00164
- Event Type
- Malfunction
- Date Received
- March 16, 2018
- Date of Event
- March 6, 2018
- Report Date
- April 5, 2018
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 00382903096466
- PMA / PMN Number
- K980987
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION: THE INITIAL MDR REPORTED THE LOT NUMBER AS UNKNOWN. THE LOT # HAS BEEN IDENTIFIED AS 8011965. D.4. LOT NUMBER: 8011965. H.6. INVESTIGATION RESULTS: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. DHR REVIEW FOR BATCH 8011965 (P/N 309646): MANUFACTURING DATES: 01/22/2018 TO 01/23/2018. BATCH QUANTITY WAS (B)(4). PRINTING AND ASSEMBLY RECORDS WERE REVIEWED AS PART OF THIS DHR REVIEW. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 8011965 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. THE REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. NO CAPA IS REQUIRED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED BY THE PHLEBOTOMIST THAT THE RUBBER STOPPER OF THE BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE ARE POPPING OFF DURING USE. NO REPORTS OF SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 187230 | BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE | SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 8011965 | 00382903096466 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |