FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM-CONTROLLER 2.0

MDR report key: 7345506 · Received March 16, 2018

Report

Report Number
3007042319-2018-01095
Event Type
Malfunction
Date Received
March 16, 2018
Date of Event
February 23, 2018
Report Date
October 30, 2018
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
UDI-DI
00888707000420
PMA / PMN Number
P100047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL IS BASED SOLELY ON THE RECEIPT OF A 3500A REPORT. SECTION F HAS BEEN UPDATED TO REFLECT THAT REPORT. F1 USER FACILITY F2 UF/IMPORTER REPORT NUMBER: 2300380000-2018-8011 F3 USER FACILITY NAME/ADDRESS: SPECTRUM HEALTH BUTTERWORTH CAMPUS 100 MICHIGAN STREET NE GRAND RAPIDS, MI 49503 F4 CONTACT PERSON: (B)(4). F5 PHONE NUMBER: 616-486-2522 F6 DATE USER FACILITY BECAME AWARE OF THE EVENT: UNKNOWN F7 TYPE OF REPORT: INITIAL F8 DATE OF THIS REPORT: APRIL 2018 F9 APPROXIMATE AGE OF DEVICE: N/A F10 EVENT PROBLEM CODES: N/A F11 REPORT SENT TO FDA: APRIL 2018 F12 LOCATION WHERE EVENT OCCURRED: HOME F13 REPORT SENT TO MANUFACTURER: YES, APRIL 2018 F14 MANUFACTURER NAME AND ADDRESS: MFR. NAME: MEDTRONIC ADDRESS: 8200 CORAL SEA STREET NE CITY: MOUNDS VIEW STATE: MN ZIP: 55112 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE CONTROLLER WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED CONTROLLER REVEALED THAT THE DEVICE PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER CONTAINED THE SMR SOFTWARE. THE SOFTWARE UPGRADE CONTAINS A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER-SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS. IN ADDITION, SEVERAL MOMENTARY DISCONNECTIONS WERE OBSERVED ON THE SECONDARY POWER SOURCE, WHICH WOULD NOT HAVE LEAD TO A POWER SWITCHING EVENT. MOMENTARY DISCONNECTIONS WILL RESULT IN AN AUDIBLE "BEEP", OR TONE. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. THE MOST LIKELY ROOT CAUSE OF THE REPORTED "BEEPING" CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CONTROLLER WAS POWER SWITCHING. THE PATIENT STATED THAT THE CONTROLLER WAS BEEPING AND FLIPS POWERS. THE CONTROLLER DOES NOT HAVE AN ALARM MESSAGE, BUT IT WILL BEEP AND THEN THE POWER SOURCE WILL FLIP TO THE OTHER ONE. THE BATTERIES ARE NOT DEPLETED AND THIS HAS HAPPENED ON ALL DIFFERENT BATTERIES. THE CONTROLLER WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
189732 HEARTWARE VENTRICULAR ASSIST SYSTEM-CONTROLLER 2.0 VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1420-CONTROLLER 00888707000420

Patients

Seq Age Sex Outcome Treatment
1 49 YR