FDA Adverse Event Malfunction Summary report: N

AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC

MDR report key: 7344707 · Received March 16, 2018

Report

Report Number
3005180920-2018-00158
Event Type
Malfunction
Date Received
March 16, 2018
Date of Event
February 8, 2018
Report Date
March 16, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 16 MARCH 2018: AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC REFERENCE LOT. 1413789 (B)(4) ITEMS MANUFACTURED AND RELEASED ON 15 DECEMBER 2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, THIS IS THE FIFTH SIMILAR EVENT REPORTED ON ITEMS OF THE SAME LOT VERSAFITCUP CC ACETABULAR SHELL CC TRIO Ø 58 REFERENCE (B)(4) (K103352) LOT 162340: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26 JULY 2016; EXPIRATION DATE: 2021-06-23. NO ANOMALY FOUND RELATED TO THE ISSUE. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

DURING THE FINAL CUP IMPACTION THE TERMINAL FOR AMIS CUP IMPACTOR REMAINED STUCK IN THE IMPLANT. THE SURGEON WAS ABLE TO REMOVE THE ASSEMBLY AND TO IMPLANT A BACK-UP SHELL OF THE SAME SIZE USING THE LINER IMPACTION SPHERE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188025 AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC MANUAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL SA 1413789

Patients

Seq Age Sex Outcome Treatment
1 Other