SPYSCOPE DS ACCESS & DELIVERY CATHETER
Report
- Report Number
- 3005099803-2018-00804
- Event Type
- Malfunction
- Date Received
- March 16, 2018
- Date of Event
- February 21, 2018
- Report Date
- February 21, 2018
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- FBN
- UDI-DI
- 08714729863236
- PMA / PMN Number
- K142922
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT'S EXACT AGE IS UNKNOWN; HOWEVER IT WAS REPORTED THAT THE PATIENT WAS OVER THE AGE OF 18. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
A VISUAL ASSESSMENT WAS PERFORMED AFTER DISINFECTION. THE CATHETER WAS KINKED IN SEVERAL LOCATIONS, BEGINNING NEAR THE DISTAL TIP. AS RECEIVED, THE WORKING CHANNEL SLEEVE PROTRUDED, CONFIRMING THE REPORTED COMPLAINT INCIDENT OF WORKING CHANNEL SLEEVE PROTRUSION. MAXIMUM WORKING CHANNEL SLEEVE PROTRUSION WAS NOTED WHEN THE SMALL KNOB WAS ARTICULATED COUNTERCLOCKWISE. A FUNCTIONAL ASSESSMENT WAS PERFORMED. THE REPORTED COMPLAINT INCIDENT OF ¿LOSS OF VISUALIZATION¿ WAS NOT CONFIRMED BECAUSE, UPON PLUGGING THE DEVICE INTO THE CONTROLLER, IT DISPLAYED A LIVE, CLEAR IMAGE. NO ISSUES WERE IDENTIFIED WITH THE IMAGE. THE DEVICE WAS FULLY ARTICULATED IN ALL DIRECTIONS; NO ISSUES WERE IDENTIFIED WITH THE IMAGE. A GUIDEWIRE WAS BACK-LOADED THROUGH THE DISTAL END OF THE SCOPE; NO ISSUES WERE IDENTIFIED WITH THE IMAGE. A SPYBITE WAS PASSED THOUGH THE WORKING CHANNEL; NO ISSUES WERE IDENTIFIED WITH THE IMAGE. THE HANDLE WAS OPENED AND IT WAS FOUND WITHIN SPECIFICATION. THE DISTAL TIP WAS CUT. THE DISTAL CAP WAS REMOVED. THE CATHETER WAS CUT OPEN USING THE CUTTING FIXTURE. THE WORKING CHANNEL SLEEVE WAS REMOVED. WITNESS MARKS WERE NOTED ON THE PEBAX. THE WHITE AREAS ALONG BOND A APPEAR TO SHOW EVIDENCE OF ADHESION. THE COMPLAINT WAS CONSISTENT WITH THE REPORTED EVENT OF WORKING CHANNEL SLEEVE PROTRUDING. BASED ON THE INVESTIGATION RESULTS, THE UNDERLYING CAUSE OF WORKING CHANNEL SLEEVE PROTRUSION IS AN INSUFFICIENT BOND, PARTICULARLY THE SECOND HEAT CYCLE OF THE WORKING CHANNEL SLEEVE BONDING PROCESS [BOND B]. THEREFORE, THE COMPLAINT INVESTIGATION CONCLUSION CODE SELECTED FOR THE WORKING CHANNEL SLEEVE PROTRUSION ISSUE IS MANUFACTURING PROCESS DESIGN. AN INVESTIGATION IS UNDERWAY TO ADDRESS THIS ISSUE. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING AN ORAL CHOLANGIOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. DURING MANIPULATION, THE SPYSCOPE DS IMAGE DISAPPEARED. THE SYSTEM WAS RESTARTED 4 TIMES, AND THE IMAGE WAS RESTORED. REPORTEDLY, NO PART OF THE SPYSCOPE DS DETACHED AND THERE WERE NO REPORTED ISSUES WITH THE ACCESSORY DEVICE. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SPYSCOPE DIGITAL ACCESS AND DELIVERY CATHETER WAS USED IN THE BILE DUCT DURING AN ORAL CHOLANGIOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE WORKING CHANNEL SLEEVE OF THE SPYSCOPE DS PROTRUDED. DURING MANIPULATION, THE SPYSCOPE DS IMAGE DISAPPEARED. THE SYSTEM WAS RESTARTED 4 TIMES, AND THE IMAGE WAS RESTORED. REPORTEDLY, NO PART OF THE SPYSCOPE DS DETACHED AND THERE WERE NO REPORTED ISSUES WITH THE ACCESSORY DEVICE. THE PROCEDURE WAS STILL COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. CORRECTION AS OF APRIL 24, 2018. THERE WERE NO ACCESSORY DEVICES USED DURING THIS PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 187738 | SPYSCOPE DS ACCESS & DELIVERY CATHETER | CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID | FBN | BOSTON SCIENTIFIC - SPENCER | M00546600 | 21643389 | 08714729863236 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |