FDA Adverse Event Malfunction Summary report: N

CYSTO-NEPHRO VIDEOSCOPE

MDR report key: 7344014 · Received March 15, 2018

Report

Report Number
2951238-2018-00192
Event Type
Malfunction
Date Received
March 15, 2018
Date of Event
February 23, 2018
Report Date
March 1, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FAJ
PMA / PMN Number
PK062049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO MAKE A CORRECTION ON THE PROCODE FROM NWB TO FAJ.

Additional Manufacturer Narrative · 1

THE CYSTOSCOPE AND THE BROKEN PORTION OF THE BENDING SECTION COVER GLUE WERE RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THAT PORTIONS OF THE BENDING SECTION COVER GLUE ON THE DISTAL END AND ON THE INSERTION SIDE WAS MISSING, THE REMAINING PORTION OF THE COVER GLUE WAS NOT DISCOLORED. THE BENDING SECTION COVER (DISTAL SHEATH) WAS CRACKED, PEELING, DISCOLORED, AND DENTED. THE INTERNAL CHANNEL OF THE CYSTOSCOPE WAS EXAMINED WITH THE USE OF A BORESCOPE AND THERE WAS NO SIGN OF MISSING PARTS OR FOREIGN OBJECT INSIDE THE CHANNEL OF THE DEVICE. THE CAUSE OF THE REPORTED INCIDENT CANNOT BE CONCLUSIVELY DETERMINED, BUT USER HANDLING CANNOT BE RULED OUT AS A CONTRIBUTORY FACTOR TO THE REPORTED INCIDENT.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT A PORTION OF THE BENDING SECTION BROKE OFF AND FELL INTO THE PATIENT'S BLADDER. THREE SMALL BLACK PIECES WAS OBSERVED IN THE PATIENT'S BLADDER AS THE CYSTOSCOPE WAS BEING WITHDRAWN FROM THE PATIENT. THE THREE BLACK PIECES OF BENDING SECTION COVER GLUE WAS RETRIEVED FROM THE PATIENT BY AN UNKNOWN MEANS. THERE WAS NO REPORTED ADVERSE IMPACT TO THE PATIENT. THE PATIENT IS REPORTEDLY DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183833 CYSTO-NEPHRO VIDEOSCOPE CYSTO-NEPHRO VIDEOSCOPE FAJ OLYMPUS MEDICAL SYSTEMS CORP. CYF-VH

Patients

Seq Age Sex Outcome Treatment
1