CYSTO-NEPHRO VIDEOSCOPE
Report
- Report Number
- 2951238-2018-00192
- Event Type
- Malfunction
- Date Received
- March 15, 2018
- Date of Event
- February 23, 2018
- Report Date
- March 1, 2019
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FAJ
- PMA / PMN Number
- PK062049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO MAKE A CORRECTION ON THE PROCODE FROM NWB TO FAJ.
THE CYSTOSCOPE AND THE BROKEN PORTION OF THE BENDING SECTION COVER GLUE WERE RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THAT PORTIONS OF THE BENDING SECTION COVER GLUE ON THE DISTAL END AND ON THE INSERTION SIDE WAS MISSING, THE REMAINING PORTION OF THE COVER GLUE WAS NOT DISCOLORED. THE BENDING SECTION COVER (DISTAL SHEATH) WAS CRACKED, PEELING, DISCOLORED, AND DENTED. THE INTERNAL CHANNEL OF THE CYSTOSCOPE WAS EXAMINED WITH THE USE OF A BORESCOPE AND THERE WAS NO SIGN OF MISSING PARTS OR FOREIGN OBJECT INSIDE THE CHANNEL OF THE DEVICE. THE CAUSE OF THE REPORTED INCIDENT CANNOT BE CONCLUSIVELY DETERMINED, BUT USER HANDLING CANNOT BE RULED OUT AS A CONTRIBUTORY FACTOR TO THE REPORTED INCIDENT.
OLYMPUS WAS INFORMED THAT A PORTION OF THE BENDING SECTION BROKE OFF AND FELL INTO THE PATIENT'S BLADDER. THREE SMALL BLACK PIECES WAS OBSERVED IN THE PATIENT'S BLADDER AS THE CYSTOSCOPE WAS BEING WITHDRAWN FROM THE PATIENT. THE THREE BLACK PIECES OF BENDING SECTION COVER GLUE WAS RETRIEVED FROM THE PATIENT BY AN UNKNOWN MEANS. THERE WAS NO REPORTED ADVERSE IMPACT TO THE PATIENT. THE PATIENT IS REPORTEDLY DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183833 | CYSTO-NEPHRO VIDEOSCOPE | CYSTO-NEPHRO VIDEOSCOPE | FAJ | OLYMPUS MEDICAL SYSTEMS CORP. | CYF-VH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |