FDA Adverse Event Malfunction Summary report: N

TC TENOTOMY SCISSDELCVD1/SERR130MM

MDR report key: 7343488 · Received March 15, 2018

Report

Report Number
9610612-2018-00100
Event Type
Malfunction
Date Received
March 15, 2018
Date of Event
August 29, 2016
Report Date
March 15, 2018
Manufacturer
AESCULAP AG
Product Code
EGN
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: THE TIP OF THE SCISSORS ARE BROKEN. PICTORIAL DOCUMENTATION WAS CARRIED OUT WITH A DIGITAL MICROSCOPE. HARDNESS MEASUREMENT WAS CARRIED OUT WITH HARDNESS TESTER. THE DEVICE HISTORY FILES HAVE BEEN CHECKED AND WERE FOUND TO BE ACCORDING TO THE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF THE INVESTIGATION THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. THIS KIND OF INSTRUMENT IS DESIGNATED FOR DELICATE USE ONLY. IT IS LIABLE THAT A MECHANICAL OVERLOAD SITUATION LED TO THE BREAKAGE. THE VISUAL INVESTIGATION AND THE HARDNESS TEST INDICATED THAT THE QUALITY REQUIREMENTS ARE IN THE SPECIFIED TOLERANCE RANGE. A CAPA IS NOT NECESSARY.

Description of Event or Problem · 0

COUNTRY OF COMPLAINT: USA. CUSTOMER STATES THAT DURING INTRA-OPERATIVE SURGERY THE TIP OF SCISSORS BROKE OFF. NO SURGICAL DELAY AND NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186005 TC TENOTOMY SCISSDELCVD1/SERR130MM DENTAL SCISSORS EGN AESCULAP AG DO250R 4507434213

Patients

Seq Age Sex Outcome Treatment
1 Other