TC TENOTOMY SCISSDELCVD1/SERR130MM
Report
- Report Number
- 9610612-2018-00100
- Event Type
- Malfunction
- Date Received
- March 15, 2018
- Date of Event
- August 29, 2016
- Report Date
- March 15, 2018
- Manufacturer
- AESCULAP AG
- Product Code
- EGN
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: THE TIP OF THE SCISSORS ARE BROKEN. PICTORIAL DOCUMENTATION WAS CARRIED OUT WITH A DIGITAL MICROSCOPE. HARDNESS MEASUREMENT WAS CARRIED OUT WITH HARDNESS TESTER. THE DEVICE HISTORY FILES HAVE BEEN CHECKED AND WERE FOUND TO BE ACCORDING TO THE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF THE INVESTIGATION THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. THIS KIND OF INSTRUMENT IS DESIGNATED FOR DELICATE USE ONLY. IT IS LIABLE THAT A MECHANICAL OVERLOAD SITUATION LED TO THE BREAKAGE. THE VISUAL INVESTIGATION AND THE HARDNESS TEST INDICATED THAT THE QUALITY REQUIREMENTS ARE IN THE SPECIFIED TOLERANCE RANGE. A CAPA IS NOT NECESSARY.
COUNTRY OF COMPLAINT: USA. CUSTOMER STATES THAT DURING INTRA-OPERATIVE SURGERY THE TIP OF SCISSORS BROKE OFF. NO SURGICAL DELAY AND NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186005 | TC TENOTOMY SCISSDELCVD1/SERR130MM | DENTAL SCISSORS | EGN | AESCULAP AG | DO250R | 4507434213 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |