FDA Adverse Event Other Summary report: N

MOBILEDART

MDR report key: 734296 · Received June 8, 2006

Report

Report Number
734296
Event Type
Other
Date Received
June 8, 2006
Date of Event
April 23, 2006
Report Date
May 17, 2006
Manufacturer
SHIMADZU MEDICAL SYSTEMS
Product Code
IZL
Report Source
User Facility report
Reporter Location
MI, US

Narratives

Description of Event or Problem · 1

THE DIGITAL PORTABLE HAS A TETHER CONNECTING THE PLATE TO THE DEVICE. THE RADIOLOGY TECHNICIAN DID NOT REMOVE THE PLATE FROM UNDERNEATH THE PATIENT PRIOR TO MOVING THE PORTABLE X-RAY EQUIPMENT AWAY FROM THE BEDSIDE. THE PLATE PULLED FROM THE BED WITH THE PATIENT STILL ON IT. THE PATIENT FELL 2-3 FEET AND LANDED FLAT ON THE LAMBS WOOL COVER AND THE X-RAY PLATE. THE PATIENT WAS ALSO INADVERTENTLY EXTUBATED UPON FALLING. WE WERE PLANNING ON EXTUBATING THE PATIENT THAT MORNING. THE PATIENT WAS BREATHING COMFORTABLY AND SKULL FILMS LOOKED FINE. HEAD CT SHOWED NO DISPLACED FRACTURE AND A SMALL RIGHT SCALP HEMATOMA IN THE RIGHT VERTEX POSTERIORLY. A POSSIBLE EQUIPMENT SOLUTION WOULD BE THAT A DRIVE CIRCUIT WOULD LIMIT THE ABILITY TO MOVE THE PORTABLE IF THE DETECTOR WAS NOT IN ITS STORAGE COMPARTMENT. ANOTHER SOLUTION MIGHT BE TO HAVE SOFTWARE THAT PREVENTS MOVING THE SYSTEM AFTER AN X-RAY UNTIL THE OPERATOR ANSWERED A QUESTION ABOUT TAKING THE PLATE OUT FROM BEHIND THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MOBILEDART X-RAY SYSTEM, PORTABLE IZL SHIMADZU MEDICAL SYSTEMS * *

Patients

Seq Age Sex Outcome Treatment
1 4 MO