FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS

MDR report key: 7338939 · Received March 14, 2018

Report

Report Number
3004209178-2018-56733
Event Type
Malfunction
Date Received
March 14, 2018
Date of Event
October 12, 2017
Report Date
March 14, 2018
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169513914
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD BLANK DISPLAY DUE TO MOISTURE DAMAGE ON LCD BOARD AND MB. UNABLE TO PERFORM IDLE CURRENT TEST, RUN CURRENT TEST, SELF TEST, OFF NO POWER TEST, DISPLACEMENT TEST AND SENSOR TEST DUE TO BLANK DISPLAY. THE PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP¿S SCREEN WAS BLANK. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 156 MG/DL. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS.