PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
MDR report key: 7338939
·
Received March 14, 2018
Report
- Report Number
- 3004209178-2018-56733
- Event Type
- Malfunction
- Date Received
- March 14, 2018
- Date of Event
- October 12, 2017
- Report Date
- March 14, 2018
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169513914
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP HAD BLANK DISPLAY DUE TO MOISTURE DAMAGE ON LCD BOARD AND MB. UNABLE TO PERFORM IDLE CURRENT TEST, RUN CURRENT TEST, SELF TEST, OFF NO POWER TEST, DISPLACEMENT TEST AND SENSOR TEST DUE TO BLANK DISPLAY. THE PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP¿S SCREEN WAS BLANK. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 156 MG/DL. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE INSULIN PUMP FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180499 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS | A5723LNASJ | 00643169513914 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |