FDA Adverse Event Malfunction Summary report: N

VERTECEM V+ CEMENT KIT

MDR report key: 7338263 · Received March 14, 2018

Report

Report Number
8030965-2018-52027
Event Type
Malfunction
Date Received
March 14, 2018
Date of Event
February 12, 2018
Report Date
February 14, 2018
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
NDN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. PATIENT INFORMATION NOT AVAILABLE FOR REPORTING. (B)(4). DEVICE MALFUNCTIONED INTRA-OPERATIVELY AND WAS NOT IMPLANTED / EXPLANTED COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. 510K#: K090432 PENDING APPROVAL. DEVICE HISTORY RECORDS REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THE: DHR REVIEW FOR: PART NUMBER: 07.702.016S, SYNTHES LOT NUMBER: 7C53190, MANUFACTURING LOCATION: (B)(4), SUPPLIER: (B)(4), RELEASE TO WAREHOUSE DATE: 14.JUN.2017. EXPIRY DATE: 01.MAR.2020. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. INVESTIGATION SUMMARY: THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. PRODUCT WAS NOT RETURNED, THEREFORE NO FURTHER INVESTIGATION IS POSSIBLE. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES ON AN EVENT IN AUSTRIA AS FOLLOWS: IT WAS REPORTED, THAT VERTECEM V+ MIXING HANDLE COULD ONLY BE MOVED WITH A GREAT AMOUNT OF FORCE. THERE WERE NO PATIENT CONSEQUENCES. THIS COMPLAINT INVOLVES 1 PART. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180986 VERTECEM V+ CEMENT KIT POLYMETHYLMETHACRYLATE BONE NDN OBERDORF SYNTHES PRODUKTIONS GMBH 7C53190

Patients

Seq Age Sex Outcome Treatment
1